GF Health Products, Inc. - www.grahamfield.com                                                                                    Liberty Service Manual 999-0822-190F AUG. 2014 

4

This product is a variable height, adjustable mattress
platform, which will provide comfort and convenience
for residents/patients and caregivers in long term
care settings.

The MAXIMUM SAFE WORKING LOAD for the
Liberty Bed with weight evenly distributed, including
bedding, resident/patient, support surface, and all
accessories, is 450 lbs. 

NEVER operate the bed if a Power Cord or Plug is
damaged or not working properly. Contact qualified
Service Personnel for examination and repair. Always
unplug the Power Cord when performing any mainte-
nance on the bed.

DO NOT open assemblies such as the Actuators, Hand
Control, or Control Box. If unauthorized personnel
perform work on these components, the manufacturer’s
warranty becomes void.

DO NOT use unauthorized parts, accessories, or
adaptors other than those specified/authorized by
GF Health Products, Inc.

When operating the High/Low, Knee, or Back Functions
of the bed, ALWAYS ensure that the confined individual
is positioned properly within the confines of the bed.
DO NOT let any extremities protrude over the side or
between the bed rails when performing these functions.

DO NOT lower the bed when objects are beneath it.
Failure to inspect under the bed can result in damage
to property or personal injury.

The bed’s Pendant Cord MUST BE ROUTED AND
SECURED PROPERLY to ensure it does not become
entangled and eventually severed during use. Also
make sure all electrical cords DO NOT get tangled
around the bed, side rails, or legs during transport or
normal operation of the bed.

When using nasal-type or masked-type administering
equipment, all oxygen or air tubing MUST BE ROUTED
AND SECURED PROPERLY to ensure that the tubing
does not become entangled and eventually severed
during the normal operation of the bed.  

Notice for California Customers -
California Proposition 65 WARNING: This prod-
uct contains a chemical known to the State of Califor-
nia to cause cancer and reproductive or developmental
harm.

The bed should ALWAYS be left in its lowest position
when unattended to reduce the risk of injury while getting
in or out of the bed.

Keep all moving parts free of obstructions (i.e. blankets/
sheets, heating blankets/pads, wiring, etc.).

DO NOT use the assist devices as push handles for moving
the bed. Assist devices can be deformed or broken if
excessive side pressure is exerted. Assist devices are not
meant for patients considered as high risks for entrapment
(i.e. patients with pre-existing conditions such as 
confusion, restlessness, lack of muscle control, altered
mental status, either organic or medicinal, or a combi-
nation thereof). Additional safety measures should be
considered for such high-risk patients. 

NEVER permit more than one (1) person on/in the bed
at any time.

Body weight should be evenly distributed over the
sleeping surface of the bed. DO NOT allow the patient
to lay, sit, or lean in such a way that their entire body
weight is placed only on the raised head or foot sections
of the bed. This especially applies when repositioning
or transferring a patient in or out of the bed. Increased
risk may occur when the patient’s size and/or weight are
inappropriate for the bed’s dimensions or weight capacity.

The bed is intended for use, storage, and transport within
a temperature range of -40˚C to +60˚C. It has a water-
resistance rating of IPX4 and IS NOT to be power-
washed or submersed. 

!

!

!

!

!

!

!

!

!

!

!

!

!

!

!

Operating Conditions

Operation of the bed is based on the following conditions:
Ambient Temperature of +10˚C to +40˚C; Relative Humidity
of 30% to 75% (Non-condensing); Atmospheric Pressure of
700hPa to 1060hPa; and a Splash Protection of IEC 60529.

Storage

Storage of the bed is based on the following conditions:
Ambient Temperature of -10˚C to 70˚C; Relative Humidity
of 10% to 100%; and an Atmospheric Pressure of 500hPa
to 1060hPa.

Radio Frequency Interference (RFI)

RFI influences most electronic equipment. Caution should be
exercised with regard to the use of portable communications
equipment in the area around such equipment. If RFI causes
erratic behavior, shut the bed off immediately. Leave it off
while the transmission is in progress. 

IMPORTANT SAFETY AND WARNING INFORMATION

LIBERTY SERIES

Содержание LBT7174

Страница 1: ...erty Service Manual 999 0822 190F AUG 2014 ETL UL CSA APPROVED UL 60601 1 IEC 60601 2 38 CSA 22 2 601 1 INSTRUCTIONS FOR USE SERVICE MANUAL To avoid personal injury or damage to bed please read all se...

Страница 2: ...the latest product information available at the time of printing The most current product information can be found online at www grahamfield com Please check all parts for shipping damage and test be...

Страница 3: ...levations 14 One Step Locking Mechanism Operation 15 Mechanical Electronic Specifications 16 IILCON Electrical Components 17 Service Part Record 18 19 Warranty Information 19 Bed Diagram with ILCON El...

Страница 4: ...ductive or developmental harm The bed should ALWAYS be left in its lowest position when unattended to reduce the risk of injury while getting in or out of the bed Keep all moving parts free of obstruc...

Страница 5: ...E NASAL OR MASK TYPE OR A TENT COVERING ONLY THE UPPER HALF HEAD END OF THE BED OXYGEN TENT CANOPIES SHOULD NOT EXTEND BELOW BED SPRING LEVEL LOCK HAND CONTROL AT FOOT OF BED WHEN USING OXYGEN ADMINIS...

Страница 6: ...es and disengages properly X Check the bottom foot pads on the foot end casters making sure they are clean of debris and undamaged Replace if needed X Check all casters to ensure that they roll proper...

Страница 7: ...lth Products Inc and Basic American Medical Products have conformed to these guidelines from a manufacturing aspect However entrapment issues can often arise when a healthcare provider facility has no...

Страница 8: ...see all actuator motor cables are routed and connected properly to the control box DISCARD 1 Cardboard Block 2 Notched Leg Foam 3 Large Blocks 4 Large Cable Tie CUT KEEP 5 Pendant Holster Pendant Powe...

Страница 9: ...identifying your bed s make and model when ordering replacement parts The Serial Number is essential if you are claiming parts or service under warranty These helpful labels can be located on the mai...

Страница 10: ...es in the brackets with the second holes from the end in the sides of the main frame rails See Detail A b For an 80 bed align the holes in the brackets with the first holes from the end in the sides o...

Страница 11: ...ress and assist devices This gap must be small enough to prevent resident patient from getting his her head or neck caught in this location STANDARD FULL WIDTH MATTRESS RETAINER INSTALLATION LIBERTY S...

Страница 12: ...head end of the bed No hardware is necessary for attaching this multi positional standard accessory Simply bend in the tabbed ends into the slots in the carriage rails and lock down into place There...

Страница 13: ...TTON 6 KNEE FOOT DECK ANGLE DOWN BUTTON OPERATION 1 To angle the Head Deck upward PRESS 1 2 To angle the Head Deck downward PRESS 2 3 To raise the Bed Up horizontally PRESS 3 4 To lower the Bed Down h...

Страница 14: ...should always be set to its lowest position when the resident patient is left unattended The head back and foot knee functions allow the resident patient to comfortably raise their head or legs In th...

Страница 15: ...ugh the slot in the top of the locking mechanism To fully lock the bed make sure the caster keepers at the head of the bed are both in their down positions Unlocked Position on Your Liberty Bed To unl...

Страница 16: ...ERTY BED MECHANICS All Models Overall Bed Length with brds wallsaver 83 86 Overall Bed Width with boards 36 00 High Height 24 50 Low Height 9 95 Maximum Head Back Deck Angle 70 Maximum Knee Foot Deck...

Страница 17: ...ty 1 Power Cable is not detachable hard wired into the side of the Control Box Head Motor Actuator Part 999 0844 052SP Quantity 1 Hi Lo Motor Actuator Part 999 0848 051SP Quantity 1 Knee Foot Motor Ac...

Страница 18: ...te Price 999 0848 051SP Hi Lo Motor Cable hard wired Hardware Pack 1 Motor 999 0844 052SP OR Head Motor Cable hard wired Hardware Pack 1 Motor 999 0844 053SP OR Foot Motor Cable hard wired Hardware Pa...

Страница 19: ...P 3 Twin Non Locking Caster with 7 16 14 Stems LBTL7174 1 Caster 999 0832 980SP One Step Locking Mechanism LBTL7174 1 Unit 100 6744 001 7 16 14 Nylon Lock Nuts secure caster stems to frame 1 Nut CAPS...

Страница 20: ...CK 999 0844 053SP Includes motor mounting hardware installation instruction sheet 4 TO ORDER A REPLACEMENT CONTROL BOX PLEASE USE 999 0848 300SP Includes Control Box and mounting clip 5 TO ORDER A REP...

Страница 21: ...and mental conditions and initiate an appropriate individual care plan to address any entrapment risks Please see page 7 for further information LIBERTY SERIES INSTALLATION 1 With the Assist Bar 1 bra...

Страница 22: ...both the deck and the deck tubes 3 From the bottom secure the screws to the tubes using four 5 16 18 Thin Nylon Lock Nuts 6 Hand tighten at this time 4 Locate the two long U shaped Rail Mounting Brack...

Страница 23: ...SERIES IMPORTANT THE MOUNTING BRACKET AND TWO CROSS BRACES MUST BE ASSEMBLED ONTO THE FOOT DECK PRIOR TO INSTALLING ANY ASSIST DEVICE TO THE FOOT END OF THE BED PLEASE ORDER THE SEPARATE KIT IN ADDITI...

Страница 24: ...ar from its vertical lock position hold onto the top of the bar with one hand and slightly pull out the black knob on the outside of the assembly with the other hand to release the locking mechanism F...

Страница 25: ...cure with the Hair Pin Clips on the other side of the Lanyard Assembly 3b 5 For Fixed Position Option 2 insert the 5 8 11 x 4 50 Cap Screws 4 from the outside through the holes in the rail and bracket...

Страница 26: ...he outside through the holes in the pivot arms and mounting bracket From the inside secure with the Hair Pin Clips on the other side of the Lanyard Assembly 3b 5 For Fixed Position Option 2 insert the...

Страница 27: ...finish is facing inward toward the head deck Align the holes in the board with the bracket holes in the Trapeze Support 2 as shown 4 From the back insert the four JCB Bolts 5 you removed in step 2 th...

Страница 28: ...4 to the inside of the footboard reusing the four bolts 5 you removed in step 2 4 Slide the new footboard assembly into the main frame tubes so that the second holes in the brackets align with the fir...

Страница 29: ...bolt to the assembly using a 5 16 18 Thin Nylon Lock Nut 4 Tighten with two 1 2 wrenches If you already have a lock mechanism system on your Liberty bed and are attaching a second Pedal Lock you are d...

Страница 30: ...TROUBLESHOOTING All Liberty Models SECTION A GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190F AUG 2014 30 LIBERTY SERIES...

Страница 31: ...Short Black Control Box Cable for Damage or Pinching c If any Cables Damaged Replace Immediately PENDANT NOT WORKING 1 Check All Pendant and Cable Connections and retest 2 Unplug Power Cord 3 Unplug t...

Страница 32: ...on Cracked Housing Scratches Bent Shaft etc 3 Head Foot Motor Works But Hi Lo Does Not a Unplug Power Cord b Remove Cord Guard at Control Box c Unplug Hi Lo Cable at the Control Box and plug into empt...

Страница 33: ...FRAMING SERVICE PARTS All Liberty Models SECTION B GF Health Products Inc www grahamfield com Liberty Service Manual 999 0822 190F AUG 2014 33 LIBERTY SERIES...

Страница 34: ...ides of bed HEAD BACK DECK 999 0822 920G KNEE DECK 999 0822 940G FOOT DECK 999 0822 930G MAIN FRAME SEAT PAN Not a Service Part LIBERTY SERIES 2 TO ORDER A REPLACEMENT KN DECK USE SERVICE PACK 999 082...

Страница 35: ...OF PAGE 26 FOR ORDER NUMBERS HEAD END Right side of Bed FOOT END Right side of Bed LIBERTY SERIES 1 00 Square Tube End Cap 100 4700 015 1 00 Square Tube End Cap 100 4700 015 1 00 Square Tube End Cap...

Страница 36: ...LOSE ANY CAPS PLUGS ON YOUR BED SEE BOTTOM OF PAGE 26 FOR ORDER NUMBERS LIBERTY SERIES 1 00 x 2 00 Rectangular Tube End Cap 100 4700 017 1 00 x 2 00 Rectangular Tube End Cap 100 4700 017 1 00 x 2 00 R...

Страница 37: ...__________________ _________________________________________________________________________________ _________________________________________________________________________________ _________________...

Страница 38: ...54937 Customer Service 1 678 291 3207 Fax 920 929 8213 www grahamfield com 2013 GF Health Products Inc All Rights Reserved Basic American Medical Products and Zenith are registered trademarks of GF He...

Отзывы: