GF Health Products, Inc. - www.grahamfield.com Liberty Service Manual 999-0822-190F AUG. 2014
26
LIBERTY SERIES
OPTIONAL HALF RAIL INSTALLATION - FOOT END
KIT ZA82900 - Includes Two Rails, Two Cross Tubes,
Two Mounting Brackets, and Hardware (Hdwr Bag 999-0820-905)
INSTALLATION
1. All Half Rails (LH & RH) mount
to the bed in a similar fashion.
Left or right designations are
based on the patient’s
orientation as he/she
lies in the bed.
2. To mount the half rails to the foot-end
of the bed, you will also need to attach
two rail brackets (#4) and two cross tubes (#5)
to the foot deck. Please see installation instructions on
page 30 of this service manual.
3. Align the holes in the rail’s pivoting arms with the holes
in the mounting brackets as shown.
4. For Quick Release (Option 1), insert the Clevis Pin ends
of the Lanyard assembly (#3a) from the outside through
the holes in the pivot arms and mounting bracket. From
the inside, secure with the Hair Pin Clips on the other
side of the Lanyard Assembly (#3b).
5. For Fixed Position (Option 2), insert the 5/8-11 x 4.50”
Cap Screws (#6) from the outside through the holes in
the rail and bracket. From the inside, secure the screws
with 5/8-11 Nylon Lock Nuts (#7)
WARNING: When assessing the Risk of Entrapment, you need to consider your bed, mattress, headboard and footboard, assist
devices (i.e. rails and assist bars) and other accessories as an entire system. All bed systems are evaluated for full compliance
to the FDA/CDRH “Hospital Bed System Dimensional and Assessment Guidance to Reduce Entrapment” guidelines.
It is extremely important to review the resident/patient’s physical and mental conditions and initiate an appropriate individual
care plan to address any entrapment risks. Please see page 7 for further information.
!
IMPORTANT:
Lubricate all pivot joints as
needed. The manufacturer
recommends Aerosol White
Lithium Grease.