-51-
Note
that the LRC for single byte commands is simply the command byte repeated. All values used in the protocol
description are represented in hexadecimal format.
Summary of Commands
Switch to Modbus Mode
Description:
Commands the Pyrometer to switch to Modbus Mode
Command Byte:
0x4D
Format:
STX 0x4D 0x4D ETX
Response:
None (pyrometer switches to Modbus Mode immediately)
Command Example:
02 4D 4D 03
Report Version
Description:
Request pyrometer version data.
Command Byte:
0x56
Format:
STX 0x56 0x56 ETX
Normal Response:
STX 0x95 <1 byte version> <9 byte PROM code> ETX
Invalid Response:
NAK
Command Example:
02 56 56 03
Example Response:
02 95 44 E2 5F 50 2B 10 10 16 73 FF 03
The 95 byte indicates that the pyrometer is running in application mode.
Version = 4.4
Prom Code = E25F502B10101673FF (factory use only)
Stop Conversions
Description:
Commands the Pyrometer to stop sending temperature readings.
Command Byte:
0x30
Format:
STX 0x30 0x30 ETX
Response:
ACK (valid), NAK (invalid)
Command Example:
02 30 30 03
Example Response :
06
Start Conversions
Description:
Commands the Pyrometer to start sending temperature readings.
Command Byte:
0x31
Format:
STX 0x31 0x31 ETX
Response:
ACK (valid), NAK (invalid)
Command Example:
02 31 31 03
Example Response:
06
Set Calibration Factor
Description:
Set the pyrometer calibration factor.
Command Byte:
0x4D
Format:
STX 0x4D <CALFACTOR> LRC ETX
Parameter Format:
4 bytes representing one 32-bit IEEE Single Precision value.
Response:
ACK (valid), NAK (invalid)
Command Example:
The following command sets the calibration factor to 0.99:
02 4D 3F 7D 70 A4 9F 03
Example Response:
06