Switch positions I to L select the power-up default filter setting of the main working elec-
trode (see table below). The selected filter frequency will be adopted at power-up, and will
remain in effect until changed via the front-panel filter controls. The factory-set default range
is 0.1 Hz. As with the power-up gain range, set the default filter to the setting you typically
use.
Power up
Dip switch position
filter
I
J
K
L
0.02 Hz
UP
UP
UP
UP
0.03 Hz
DN
UP
UP
UP
0.04 Hz
UP
DN
UP
UP
0.05 Hz
DN
DN
UP
UP
0.06 Hz
UP
UP
DN
UP
0.08 Hz
DN
UP
DN
UP
0.10 Hz
UP
DN
DN
UP
0.15 Hz
DN
DN
DN
UP
0.20 Hz
UP
UP
UP
DN
0.30 Hz
DN
UP
UP
DN
0.40 Hz
UP
DN
UP
DN
0.50 Hz
DN
DN
UP
DN
0.80 Hz
UP
UP
DN
DN
1.00 Hz
DN
UP
DN
DN
Switch positions M, N, and O select the PC-REMOTE selection (see table below). These set-
tings will allow up to 8 detectors to be independently addressed by a remote computer. Each
independent detector must be given a unique identification by setting dip switches M, N and
O. The default setting is W1.
Remote
Dip switch position
address
M
N
O
W1
UP
UP
UP
W2
DN
UP
UP
W3
UP
DN
UP
W4
DN
DN
UP
W5
UP
UP
DN
W6
DN
UP
DN
W7
UP
DN
DN
W8
DN
DN
DN
Switch position P enables/disables PC remote control. The default setting is Remote En-
abled.
PC remote
P
Remote enabled
UP
Remote disabled
DN
NOTE: Enabling PC remote does not put the LC-4C in REMOTE, but allows the PC interface
to control REMOTE/LOCAL operation.