BCS36x8
ex
Series
Connection example
Type 17-A1S4-*HP* and B7-A2S4-****
Subject to technical changes.
05/2020
Page 43 of 78
9.1.4
Testing communication/data transmission
The communication/data transmission on the host PC can be tested with the help of a terminal program or a
software keyboard wedge.
A software wedge or other application must be installed on the host PC for data transmission
when using a serial interface.
The software wedge or other application is used to convert the incoming data to a keyboard
entry and to enter the data into the field currently active on the host PC.
Terminal program
The test with a terminal program described below was conducted with the H-Term application.
Alternatively any other terminal program that supports serial communication may be used.
Testing communication/data transmission
Terminal Programm öffnen
Select serial interface to which the BCS3608
ex
is connected (COM1 in the example).
Can be checked in the device manager.
Perform settings in the terminal program and press ‘Connect’.
Read barcode with scanner and check in the terminal program that the data are displayed.
Data transmission is OK if the barcode data is displayed in the “Received Data” window.
If not, check connection and programming.