Maintenance
70
Operating Instructions HYGROPHIL® H 4230-12 BA 060207
Push the measuring cell downwards and out of the middle measuring
chamber section. Attention: never press the “tulip” but press carefully
beside the “tulip” to the measuring cell by using a screw driver or a simi-
lar tool.
6.3.5
Changing the humid temperature sen-
sor
Detach the knurled nut from the sensor mount and carefully extract the
sensor.
Insert the new sensor and tighten the knurled nut.
After reassembling the measuring chamber, connect the sensor’s con-
nector in the enclosure and tighten the knurled nut.
Sensor calibration see section 5.2.4.8.