PROFIBUS
98
Operating Instructions HYGROPHIL® H 4230-12 BA 060207
DaAdr $3B: Fullscale
Format: Array [1...2] of Float
Before you can write this variable you must first change to standby mode.
Setpoint value that corresponds to 100% in the output.
Default:
Setpoint [x]
Designation/unit
Value
1
DT/°C
100
2
HT/°C
100
DaAdr $3C: Zeropoint
Format: Array [1...2] of Float
Before you can write this variable you must first change to standby mode.
Setpoint value that corresponds to 0% in the output.
Default:
Zeropoint [x]
Designation/unit
Value
1
DT/°C
0
2
HT/°C
0
8.3.5
1-channel parameters
DaAdr(hex)
Identifier
Memory-Type
Read-out
Format
$40
Heating_w
EERPOM read/write
dec
Float
$41
Heating_xp
EERPOM read/write
dec
Float
$42
HeatingControlStatus
RAM read/write
-----
Record
$46
ErrorTime
EEPROM read/write
dec
Integer
$47
HT ohne Komp
RAM read
dec
Float
$48
T2KorrektConfig
EEPROM read/write
dec
Char
$4C
Timeout Aktivierung
EEPROM read/write
hex
Char
DaAdr $40: Heating_w
Format: Float
Before you can write this variable you must first change to standby mode.
Setpoint value w for the hose heating control (actual value =TT).
Default = 120°C
DaAdr $41: Heating_xp
Format: Float
Before you can write this variable you must first change to standby mode.
Proportional range xp in %.
Default = 3%
Содержание HYGROPHIL H 4230-12 Ex Zone II
Страница 2: ......
Страница 30: ...System description 24 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Страница 34: ...Safety precautions 28 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Страница 66: ...Programming 60 Operating Instructions HYGROPHIL H 4230 12 BA 060207...
Страница 112: ...Operating Instructions HYGROPHIL H 4230 12 BA 060207...