32
T
ROUBLESHOOTING
Problem (cont.)
Possible Causes (cont.)
Action (cont.)
0.2 micron final filter clogs
Possible feedwater contamination.
Call Technical Service for possible
rapidly after replacement.
feedwater testing.
Cartridges not properly rinsed up
Rinse up cartridges as described in
before use.
Cartridge Rinse-Up Procedures
.
Replace the 0.2 micron final filter
assembly as indicated in the
0.2
Micron Final Filter Replacement
section.
EASYpure II UV contaminated
Sanitize EASYpure II UV according
with bacteria.
to the instructions in
System
Sanitization.
Replace the 0.2 micron
final filter assembly as indicated in the
0.2 Micron Final Filter Replacement
section.
Short cartridge life.
Cartridges being used are beyond
Check the expiration date.
expiration date.
Cartridges begin to lose capacity
after being stored two years from
the date of manufacture. Replace
the cartridges with unexpired ones.
Change in feedwater characteristics.
Call Technical Service for possible
feedwater testing.
Increased product water usage.
Verify usage.
Water leakage inside EASYpure II UV.
Loose connections.
Tighten connections.
Tubing is not inserted completely.
Insert tubing completely. See
Push-to-Connect Tubing
Installation
section of this manual.
Missing or defective cartridge
Install or replace cartridge o-rings.
o-rings.
Leak at cartridge.
Make sure cartridge access door is
closed and latched.
Water leakage at a final filter.
Not installed far enough.
Install or screw in further and/or
add Teflon tape.