Black plastic cover
Receptacle
UV lamp
UV chamber
I
NSTALLATION
13
Caution
Do not touch the glass portion of the
UV lamp. We recommend that you
wear lint-free gloves when handling the
lamp. The glass portion must be free of
fingerprints, perspiration, etc. Even a
single fingerprint will reduce the effec-
tiveness of the lamp. If you accidentally
touch the glass portion of the lamp,
clean the lamp with a lint-free cloth;
use isopropyl alcohol if required.
Caution
Do not rotate the metal cover as this
will loosen the water tight seal and
thus damage the UV lamp.
Note
If UV lamp is not installed properly
an “Er3” message will appear for 15
seconds of every minute.
Component Installation
UV Lamp Installation (See Fig. 5)
Before you mount the EASYpure II UV, install the UV lamp as
follows:
1.
Remove the screws securing the EASYpure II UV
top cover.
2.
Remove the top cover by lifting straight up.
3.
Locate the UV oxidation chamber and pull the top
black plastic cover off.
Do not pull on the cable.
4.
Remove the lamp from its container.
DO NOT
TOUCH THE GLASS PORTION OF THE LAMP.
It
is recommended that lint free gloves be worn when
handling the lamp. The glass portion must be free
of fingerprints, perspiration, etc. Even a light coat-
ing of perspiration will reduce the effectiveness of
the lamp.
5.
Clean the lamp with isopropyl alcohol and a lint
free cloth.
6.
Carefully insert and hold the UV lamp partially into
the UV chamber.
7.
Connect the UV lamp to the receptacle in the black
plastic cover. Slide the lamp completely into cham-
ber and replace the black plastic cover on the UV
chamber.
8.
Reinstall the EASYpure II UV top cover.
Figure 5: UV Lamp Installation
Warning
This unit is equipped with an
ultraviolet lamp. Ultraviolet radiation
is harmful to the eyes and skin. Do
not observe the lamp directly.
Note
The UV lamp contains mercury. If
broken or no longer needed, do not
dispose of the UV lamp in the trash.
Recycle or dispose of the UV lamp as
hazardous waste.