19
Switches
When the main power switch (on the lower right side of the
unit) is on, the 4 switches on the control panel function as
follows.
FLUSH
(followed by ENTER/RESTART) places the unit
into a 4-minute flush.
MEMBRANE
(followed by ENTER/RESTART) resets the
membrane operational timer to zero.
CARBON
(followed by ENTER/RESTART) resets the car-
bon operational timer to zero.
ENTER/RESTART
is used to select an operation, or
restart the unit if it is in standby mode and reservoir high
LED is not illuminated. This key is also used to restart the
unit if it has shut down due to an extended low membrane
pressure condition.
Electrical/Reservoir Connections
DIAMOND RESERVOIR CONNECTION
Utilized with 30 liter reservoir - D12681
REMOTE DISPLAY
Utilized when control panel is installed in a remote loca-
tion.
ALTERNATE RESERVOIR CONNECTION
Black jumper plug must remain in place unless either a 60
liter, 100 liter or other Barnstead reservoir is being
installed.
C
ONTROLS
DIamond
Reservoir
connection
Remote Display
Alternate
Reservoir
connection
Power Switch
Cable Connections
Location
Note
The black protective cover on the remote
display connector should remain in place
when the display is not remotely located.
Power Switch: | = ON