33
SPECIFICATIONS
Display:
Two lines of 16 characters. Four-digit
process value. Four-digit setpoint value.
Accuracy:
Type J,K,T,E and N
above
−
100°C (
−
148°F): ±0.1% reading, ±0.4°C (±0.7°F)
below
−
100°C (
−
148°F): ±0.1% reading, ±1°C (±1.8°F)
Type R,S and B
±0.1% reading, ±1°C (±1.8°F)
Thermistor, RTD
±0.1% reading, ±0.4°C (±0.7°F)
Resolution:
0.1° auto-ranging to 1° above 999.9
Environment:
Temperature, Operating:
0 to 40°C
Temperature, Storage:
−
20 to 70°C
Temperature, Specification:
18 to 28°C
Humidity:
10 to 90% RH (non-condensing)
Altitude:
less than 2000 m
Pollution Degree:
Pollution degree 2 per IEC 664
(Indoor Usage—lab, office)
Warm-Up time:
10 minutes
Construction:
Dimensions:
W:7.3 in H:3.7 in D:10.0 in
Weight:
5.3 lbs
Enclosure:
IP52 per IEC 529
Compliance:
115V and 230V:
UL 916
230V (For CE Mark):
EN61010-1/A2:1995 and
EN61010-2-010:1995
(EU Low Voltage Directive) and
EN61326-1/A1:1998
(EU EMC Directive)
Sensor Input Range/Accuracy:
Thermocouples
Range
Type
E:
−
200 to 1000°C
(
−
328 to 1832°F)
J:
−
190 to 1000°C
(
−
310 to 1832°F)
K:
−
200 to 1372°C
(
−
328 to 2502°F)
N:
−
200 to 1300°C
(
−
328 to 2372°F)
T:
−
200 to 400°C
(
−
328 to 752°F)
B:
200 to 1800°C
( 392 to 3272°F)
R:
0 to 1768°C
(
32 to 3214°F)
S:
0 to 1768°C
(
32 to 3214°F)