Operation
8
Barkey GmbH & Co. KG - Gewerbestrasse 8 - 33818 Leopoldshoehe - Germany - Head Office: +49 (5202) 9801-0
Customer Service: +49 (5202) 9801-30 Fax: +49 (5202) 9801-99 - e-mail: [email protected]
Service manual
Barkey warming center III DTM
4
Operation
4.1
Display elements.
Display elements are located on the front of the Barkey warming center III DTM. They are
integrated in the front face of the control module. The following illustration shows an
overview of the display elements.
Operating display
Fault display
Heating activity display
Temperature display
Figure 2: Display panel
The following subsections explain the meaning of individual display elements.
4.1.1
Operating display
The operating display indicates the heating device operating status.
Status
Meaning
Off
Power supply is not available.
The heating device is not switched on.
On
Power supply is available.
The heating device is switched on and active.
Flashing
Power supply is available.
The heating device is switched on.
The door is open.
The circulating air temperature regulation is switched off.
An acoustic signal will sound after two minutes
Signal to remind you that the door is open.