12
4.
Installation
and starting the device
Before starting the device, check the completeness of the
set and make sure if it has no defects or mechanical dam-
age. In case of any noticed damage, contact the distribu-
tor or manufacturer of the device. Do not proceed with the
installation and starting the machine if the packaging is
damaged. Unpack the device and place it on a stable sur-
face adequate for its weight. Remember not to place the
device too close to the wall, because the holes of cooling
fans are located on the back of the case. Do not place on
or near the device other electrical devices that may gener-
ate electromagnetic field or high temperature and objects
containing water or other liquids.
When the device is delivered in autumn or winter months,
when the ambient temperature is below 10ºC, wait about
2 hours before starting the device to equilibrate the tem-
perature of the room and the device. The condensation of
water vapor may cause an electric shock or fire.
Содержание CarePump Lite 6
Страница 1: ...UM_EN_CPLITE6_04 Lymph Drainage Therapy Device Pressotherapy Device User Manual ...
Страница 3: ...3 9 Warnings 27 10 Troubleshooting 28 11 Maintenance and storage 30 12 Utilization 32 ...
Страница 25: ...25 The basic CarePump set includes Central unit device Power cord 7 Contents of the package A B A B ...
Страница 34: ...Manufacturer BardoMed Sp z o o ul Konecznego 6 66 31 216 Kraków NIP 945217693 ...