ALC & DICOM O
PTIONS
SUBMENU
*
*This menu is available only in MDNC-2121, MDNC-3121, MDNG-2121
and MDNG-6121.
Correction value
Shows the ambient light correction value
that is taken into account in the calculation
of the display function, measured in lux.
DICOM offset
The DICOM DF is recalculated at the
moment the DICOM Offset is changed,
taking the new DICOM Offset into account.
•
When set to “Dark Room”, the
ambient light is not taken into
account.
•
When set to “Preset”, a preset
ambient light value determined by
the selected reading room is taken
into account.
Reading room
Allows to select from a pre-defined list of
reading room types.
You must select a reading room that
corresponds to the type of room the
display is installed in.
Reading room def.
Jumps to the reading room definition
submenu, which allows to edit the reading
room condition settings.
Calibration info
Jumps to the calculation information
submenu, which displays information
about the values taken into account to
recalculate the DICOM DF.