background image

Troubleshooting

(This page intentionally left blank.)

102

Nio MKII

(This page intentionally left blank.)

Содержание Nio MKII

Страница 1: ...Installation User Manual Nio MKII Supported displays E 2621 E 3620 MDNG 5121...

Страница 2: ...This page intentionally left blank 2 Nio MKII This page intentionally left blank...

Страница 3: ...ved Nor this document nor any part of it may be reproduced or copied in any form or by any means graphical electronic or mechanical including photocopying taping or information storage and retrieval s...

Страница 4: ...4 Nio MKII This page intentionally left blank...

Страница 5: ...oller installation 32 BarcoMed PCI Express Controllers 35 Display installation 37 After unpacking the display 37 Power connection 40 Video connection 41 USB connection 42 Cable routing 42 Attaching th...

Страница 6: ...operation 71 Display settings 74 General 74 Display tab 75 Graphic Board tab 76 Calibration tab 77 Test patterns 82 Application settings 83 Calibration tab 83 MediCal Administrator tab 84 Update NioWa...

Страница 7: ...Driver re installation updates or removal 109 Reinstalling or updating your BarcoMed NIO driver 109 Uninstalling the BarcoMed NIO driver or Barco NioWatch software 110 Technical Information 115 Techn...

Страница 8: ...Table of Contents This page intentionally left blank 8 Nio MKII...

Страница 9: ...Nio MKII 9 Preface...

Страница 10: ...Preface This page intentionally left blank 10 Nio MKII This page intentionally left blank...

Страница 11: ...is device complies with Part 15 of the FCC Rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference re...

Страница 12: ...at to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Canadian notice This ISM device complies with Canadian ICES 001 Cet appareil ISM est conforme la...

Страница 13: ...with external power supply Class III equipment Degree of safety flammable anesthetic mixture Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with oxygen o...

Страница 14: ...o cords at plugs and receptacles Water and moisture Never expose the display to rain or moisture Never use the display near water e g near a bathtub washbasin swimming pool kitchen sink laundry tub or...

Страница 15: ...and activating a good screen saver may decrease the risk of image sticking image retention This apparatus conforms to CE0120 MDD 93 42 EEC class IIb product IEC 60601 1 UL 60601 1 CAN CSA C22 2 No 60...

Страница 16: ...ates the display is approved according to the UL regulations or Indicates the display is approved according to the c UL regulations Indicates the display is approved according to the DEMKO regulations...

Страница 17: ...s Indicates this apparatus must not be thrown in the trash but must be recycled according to the European WEEE Waste Electrical and Electronic Equipment directive Warning Risk of injury to human being...

Страница 18: ...Explanation of symbols This page intentionally left blank 18 Nio MKII...

Страница 19: ...Nio MKII 19 Overview...

Страница 20: ...Overview This page intentionally left blank 20 Nio MKII This page intentionally left blank...

Страница 21: ...3620 is a 20 8 inch grayscale LCD display with a native resolution of 2048 x 1536 The MDNG 5621 is a 21 3 inch grayscale LCD display with a native resolution of 2560 x 2048 It is available in a clear...

Страница 22: ...play will power down the display The power saving system can be switched on or off using the on screen menus Tilt swivel base The versatile tilt swivel foot allows to use the display for viewing portr...

Страница 23: ...version Windows XP Professional Windows 2000 Professional Windows Server 2003 Windows XP Professional x64 Edition Windows Server 2003 x64 Edition Features of the BarcoMed NIO display controller Portr...

Страница 24: ...le on head 1 when the OS is booted in VGA mode 640x480 Default Refresh Rate 16 colors 4 bit color grayscale 800x600 Default Refresh Rate 256 colors 8 bit color grayscale System Configuration Guideline...

Страница 25: ...re and documentation Display box Each display box includes one display and a display accessory box containing the following items Plastic cover of the tilt swivel foot Power supply Digital video DVI c...

Страница 26: ...screen display OSD The LED is green when the display is on when enabled in the on screen menus The LED is orange when the display is in Stand by power saving mode 2 USB downstream port See also item...

Страница 27: ...x when the display is shipped to the customer 3 Tilt swivel foot 4 Slot for security cable e g Kensington lock 5 DVI digital video input 6 USB upstream port Connect this connector to the PC USB bus if...

Страница 28: ...s connector 9 Tilt swivel foot clip The display is shipped with this clip in the foot to protect the tilt swivel mechanism during transport After unpacking you should remove this clip Do not throw the...

Страница 29: ...Nio MKII 29 Installation...

Страница 30: ...Installation This page intentionally left blank 30 Nio MKII This page intentionally left blank...

Страница 31: ...ection during transport Avoid reflections in the flat panel to reduce eye strain Place the display on a strong and stable table or desk Keep the display away from heat sources and provide enough venti...

Страница 32: ...in your PC please take a few minutes to familiarize yourself with both the display controller s and the PCI slots in your computer Figure 3 The BarcoMed NIO display controller Using the VGA capabilit...

Страница 33: ...host with VGA enabled you should enable VGA on only ONE of the BarcoMed NIO display controllers and disable VGA on ALL the other BarcoMed NIO display controllers Examples of PCI slots Although the Ba...

Страница 34: ...ies of the BarcoMed NIO see Using the VGA capabilities of the BarcoMed NIO display controller on page 32 you must now remove any VGA display controller s that are currently installed in the computer o...

Страница 35: ...of the displays may vary when you are running multiple BarcoMed NIO display controllers This is due to the PC s PCI bus control in the system BIOS and not the BarcoMed display controller It may becom...

Страница 36: ...ion in PCI Express x8 or x16 slots Optimum performance will always be obtained by installing them in a PCIe x8 slot Depending on your PC manufacturer s implementation of the logic for the PCIe x16 slo...

Страница 37: ...display foot is blocked with a red clip to prevent damage during transportation Before installing the display you must remove this clip Figure 8 To remove the clip 1 Position the display with its rea...

Страница 38: ...fore connecting the cables To change the panel orientation 1 Stand at the front side of the panel and take the panel at both sides 2 Very important Tilt the panel before changing the orientation Shoul...

Страница 39: ...ape turn the panel counterclockwise Figure 12 To rotate the panel from portrait to landscape 4 To change from landscape to portrait turn clockwise Notice If after installing the display or the system...

Страница 40: ...y controller board the image orientation will not change with the panel orientation To change the orientation of the image you will have to change the resolution in the Windows Display control panel i...

Страница 41: ...necting DVI cables One display Figure 15 1 Connect one end of the DVI cable to the DVI input of the display 2 Connect the other end of the DVI cable to the DVI connector of the display controller boar...

Страница 42: ...s USB upstream connector by means of a USB cable 2 Connect any USB device to one of the display s USB downstream connectors Cable routing Routing the signal cables Bind the cables in the connector com...

Страница 43: ...y box At last put the foot cover back in place To put the foot cover in place Figure 17 1 Push the upper side of the cover onto the foot so that the hooks inside the cover are positioned right under t...

Страница 44: ...pproved by VESA according to the VESA 100 mm standard Use an arm that can support a weight of at least 13 kg 28 66 lbs To attach the display to an arm stand 1 Put the display face down on a clean and...

Страница 45: ...ing the display to an arm stand Nio MKII 45 6 Remove the four screws fixing the foot while supporting the foot 7 Attach the arm stand firmly to the panel using 4 screws M4 x 8 mm Figure 19 4 screws M4...

Страница 46: ...el s to the BarcoMed NIO display controller s and power supply s 3 Decide if you want to install the NioWatch Software in addition to the driver 4 Decide if you are going to install the Nio driver wit...

Страница 47: ...lease you may skip steps 2 and 3 below 1 Boot your system and log in using an account with administrator privileges 2 When the Found New Hardware Wizard comes up click Cancel When the System Settings...

Страница 48: ...ure 20 appears you may either click Next to accept the default settings or if you know the settings required for your viewing application you may select them now and then click Next You may change the...

Страница 49: ...detect your Barco displays or fails to attach them to the desktop correctly please refer to the section Setting the resolution of your Nio display on page 105 in the Troubleshooting section of this ma...

Страница 50: ...License Agreement on the next screen click Yes to continue 3 Click Next on the Choose Destination Location screen to install the NioWatch software in the default location Or click Browse to install t...

Страница 51: ...r intervention is required during installation To install NioWatch in silent mode 1 Open the command prompt DOS window in Windows 2 Using DOS commands navigate to the folder containing the NioWatch se...

Страница 52: ...NioWatch This page intentionally left blank 52 Nio MKII...

Страница 53: ...Nio MKII 53 Display Controller settings...

Страница 54: ...Display Controller settings This page intentionally left blank 54 Nio MKII This page intentionally left blank...

Страница 55: ...correct region via the Regional Settings Control Panel in your machine s Start Settings Control Panel Using the BarcoMed Driver Tab Please note that you must have logged on to Windows using an account...

Страница 56: ...e Status section displays information about the current BarcoMed display controller driver and the currently selected display resolution Graphics Board This displays the current BarcoMed display contr...

Страница 57: ...bit 256 color modes should continue to work normally Please note that dithering is not used while in this mode The Enable Dithering check box will be grayed out and dithering will be automatically di...

Страница 58: ...ing gray values so that the palette has the full 256 gray values within it Figure 24 Please note that dithering is not permitted while in this mode The Enable Dithering check box will be grayed out an...

Страница 59: ...ble Dithering check box will be grayed out and dithering will be automatically disabled regardless of whether this check box is checked This complies with the Windows standard interface method If you...

Страница 60: ...stance these colors seem to blend into a single intermediate color The technique is similar to the half toning used in black and white publications to achieve shades of gray Please note that this opti...

Страница 61: ...out the Firmware installed on the board Table 1 AURA Controllers Non AURA Controllers BarcoMed Hardware Tab Support BarcoMed Nio PCI and Firmware Information BarcoMed Coronis PCI and Firmware Informat...

Страница 62: ...rcoMed 3MP2FH PCI and Firmware Information BarcoMed 5MP1H PCI Information Only BarcoMed 5MP1HM PCI and Firmware Information BarcoMed 5MP2 PCI Information Only BarcoMed 5MP2 AURA PCI and Firmware Infor...

Страница 63: ...BarcoMed display controller driver and the currently selected display resolution Identify Device This button is for BarcoView Medical Imaging Systems MIS internal use only and is grayed out PCI Infor...

Страница 64: ...ed PCI slot Hardware Version Displays the Hardware Version for the BarcoMed display controller installed in the selected PCI slot Advanced Button By clicking on this button the user can display more i...

Страница 65: ...ese1 The application will detect the regional settings on the user s machine and switch languages accordingly The default language is English Welcome Screen The Welcome screen reminds the user to clos...

Страница 66: ...ations data This screen gathers the diagnostic data from the user s system as described above in this document Click the Start Diagnostic button to begin the diagnostic gathering process To perform th...

Страница 67: ...xam has collected the data the user can submit it to ImageCare Barco Medical Imaging Systems customer support organization by email To do this you need a system with access to the World Wide Web 1 Ent...

Страница 68: ...is button will launch the BarcoMed Hardware Configuration utility This program allows the user to flash update the firmware stored in the ROM of the currently selected BarcoMed display controller The...

Страница 69: ...rmat which guides the user through the flash update procedure The user will be prompted to select a firmware update file to use for the update process This file will be provided by BarcoView MIS if an...

Страница 70: ...BarcoMed Hardware Tab This page intentionally left blank 70 Nio MKII...

Страница 71: ...Nio MKII 71 NioWatch operation...

Страница 72: ...NioWatch operation This page intentionally left blank 72 Nio MKII This page intentionally left blank...

Страница 73: ...the NioWatch console appears allowing you to execute the NioWatch functions Figure 28 NioWatch console Display settings Allows to view information about your dis play s and display controller Also al...

Страница 74: ...in the Display drop down box If you wish to control a display in a multi head system you must select the display controller head to which the display is connected You can do this by clicking on the c...

Страница 75: ...calibrated to the selected display function in the past the display will be set according to this calibration If no former calibration was found default factory settings are selected When you restart...

Страница 76: ...DICOM compliant you must select the DICOM display function and calibrate the display After calibration the luminance must remain in calibrated position to maintain DICOM compliance Graphic Board tab F...

Страница 77: ...o calibrate the display using the Barco LCD sensor 1 If you have multiple displays of the same type connected multi head system consider if you wish to equalize the display you calibrate to a referenc...

Страница 78: ...brate button 4 A message appears showing the different calibration options Figure 32 Calibration choices 5 Select Using the BARCO LCD sensor and click OK This option is grayed in case NioWatch does no...

Страница 79: ...he display number that appears in this option refers to the reference display Manual calibration 1 Follow step 2 and 3 of the calibration with Barco LCD sensor see page 77 2 In the Calibration dialog...

Страница 80: ...peat this procedure until the button is grayed This indicates the last point is set 6 Click OK to finish To reset to the latest calibration 1 Follow step 2 and 3 of the calibration with Barco LCD sens...

Страница 81: ...calibration 1 In the Calibration tab dialog click on Show calibration chart This button is available only if the display has been calibrated with a sensor before 2 A graph is shown displaying the lat...

Страница 82: ...ern by clicking on the corresponding thumbnail or selecting a pattern by name from the Test pattern drop down box 2 Click Show to display the pattern If you have selected Custom Image you can select a...

Страница 83: ...ame type and you wish display 2 to have the same output luminance as display 1 Therefore first calibrate display 1 Next calibrate display 2 while the Equalization option is checked The first calibrate...

Страница 84: ...l be calibrated and additionally the display luminance will be matched to the luminance of the reference display 1 MediCal Administrator tab Figure 38 MediCal Administrator tab Here you can connect to...

Страница 85: ...r friendly web interface accessible from any client users have access to all the information of the installed display base For more information about MediCal Administrator please contact Barco Medical...

Страница 86: ...e NioWatch application is installed during installation of NioWatch To update NioWatch Select Update NioWatch from the Start Programs Barco NioWatch menu The application will search via the Internet i...

Страница 87: ...Nio MKII 87 Display operation...

Страница 88: ...Display operation This page intentionally left blank 88 Nio MKII This page intentionally left blank...

Страница 89: ...ommended Switching on DPMS on display and PC and activating a good screen saver may decrease the risk of image sticking image retention Stand by switching When the display is on and no on screen displ...

Страница 90: ...When the OSD is not on the screen rotate the control wheel to display the OSD The main menu appears 2 To enter into a menu First rotate the control wheel to select the desired menu Next press the whee...

Страница 91: ...value from a predefined list e g in Settings rotate the control wheel to select the setting and press the control wheel until the desired value appears Locking and unlocking user controls The User Co...

Страница 92: ...e 3 Press the control wheel 2 times 4 Rotate the control wheel 1 step counterclockwise The on screen display appears Note Steps 2 to 4 must be performed in maximum 3 seconds 5 Rotate the control wheel...

Страница 93: ...t affect the grayscales of the image on the screen Name Description Luminance Adjust the target luminance to which the dis play will be stabilized Settings Change settings for DPMS Power LED and User...

Страница 94: ...ode the LED will turn orange even if it was switched off by this setting User Controls Disable the control wheel functions Name Description Product The display type Serial No Indicates the display ser...

Страница 95: ...Nio MKII 95 Cleaning...

Страница 96: ...Cleaning This page intentionally left blank 96 Nio MKII This page intentionally left blank...

Страница 97: ...ss panel or cabinet as excess liquid may cause damage to internal electronics Instead apply the liquid to the cleaning cloth DO NOT USE Lye or cleaning solutions containing lye Acid Detergents with fl...

Страница 98: ...is standard installed on E 2621 and E 3621 displays It is optional for other Nio display models Cleaning instructions Clean the glass using a soft cotton cloth lightly moistened with a watery solution...

Страница 99: ...aminants on the surface of the lcd panel or the glass panel Attention The dust remover contains a liquid gas If you shake the can or move the can too fast while spraying you may blow drops of liquid o...

Страница 100: ...oth lightly moistened with a recognized cleaning product for medical equipment Repeat with water only Wipe dry with a dry cloth The cabinet has been tested for resistance to the following products Cid...

Страница 101: ...Nio MKII 101 Troubleshooting...

Страница 102: ...Troubleshooting This page intentionally left blank 102 Nio MKII This page intentionally left blank...

Страница 103: ...103 General tips If one display from a multi head system exhibits problems try to eliminate the problem by switching video cables or power supplies In that way you can find out if the problem resides...

Страница 104: ...s noise or interference The video cable may be of poor quality DVI video cable may not be firmly connected to the PC or to the display Image contains missing pixels A number of missing pixels may be n...

Страница 105: ...rotated 90 Select another resolution in the Windows Display Properties control panel The previous image remains slightly visible on the screen This phenomenon called image sticking is normal if the s...

Страница 106: ...this time If you installed your BARCO NIODisplay System drivers in SingleView mode default for Windows 2000 there will be one rectangle for the virtual display representing the two heads controlled b...

Страница 107: ...o display should now synchronize and display the Windows desktop 8 Click Yes when asked Your desktop has been reconfigured Do you want to keep these settings To set the resolution of the second displa...

Страница 108: ...General tips 108 Nio MKII arrangement in the window on the Settings tab matches the physical arrangement of your configuration...

Страница 109: ...for you To reinstall or update only the BarcoMed NIO driver follow the steps described in the section Using the BarcoMed Product Installation Wizard with the following changes 1 Boot your system and l...

Страница 110: ...inished installing click Finish Then click Finish again Reboot the system when prompted and then reset the resolution of your displays if necessary Uninstalling the BarcoMed NIO driver or Barco NioWat...

Страница 111: ...izard 2 The BarcoMed Product Installation Wizard will display its welcome screen 3 Make certain that the checkbox next to the driver is checked and that all other checkboxes are unchecked figure 46 be...

Страница 112: ...l process Click Finish again to exit the wizard Click Yes if Windows tells you that you must restart your computer before the new setting will take effect 8 If you have installed your driver in DualVi...

Страница 113: ...estart Finish uninstalling the driver by clicking Next three times Then click Reboot Finish and Yes If the BarcoMed Product Installation Wizard doesn t automatically restart finish uninstalling the dr...

Страница 114: ...Driver re installation updates or removal This page intentionally left blank 114 Nio MKII...

Страница 115: ...Nio MKII 115 Technical Information...

Страница 116: ...Technical Information This page intentionally left blank 116 Nio MKII This page intentionally left blank...

Страница 117: ...mm H x 0 27 mm V Native color resolution 8 bits sub pixel Luminance 500 cd m calibrated Contrast ratio 700 1 on off in dark environment Response time 17 5 ms typical 25 C after 30 min warm up Controls...

Страница 118: ...585 x 250 mm Net weight 14 kg Operating Temperature Operation 0 C to 40 C Within specs 15 C to 35 C Storage Temperature 20 C to 60 C Humidity 8 80 non condensing for opera tion 5 95 non condensing fo...

Страница 119: ...o DVI Rev 1 0 specifications USB standard supported USB 1 1 Power source Input for 12 VDC power supply unit 90 264 VAC Input for display 12 VDC The supplied 12VDC power supply must be used Power consu...

Страница 120: ...ast ratio 800 1 on off in dark environment Response time 25 ms typical 25 C after 30 min warm up Controls Push turn control wheel for stand by switching and OSD controls Input connectors DVI dual chan...

Страница 121: ...ertical position high est position tilt 0 swivel 0 408 x 591 x 250 mm Net weight 11 1 kg Operating Temperature 0 C to 40 C 15 C to 35 C within specs Storage Temperature 20 C to 60 C Humidity 8 80 non...

Страница 122: ...in layout Pin no Signal Pin no Signal 1 TMDS DATA 2 14 5V POWER 2 TMDS DATA 2 15 GND 3 GND 16 HOT PLUG DETECT 4 NC 17 TMDS DATA 0 5 NC 18 TMDS DATA 0 6 DDC CLOCK 19 GND 7 DDC DATA 20 NC 8 NC 21 NC 9 T...

Страница 123: ...some high end boards like some of the BarcoMed boards contain two sets of video and sync outputs This is called a dual head display controller It is like two complete display controllers implemented...

Страница 124: ...Glossary This page intentionally left blank 124 Nio MKII...

Страница 125: ...Nio MKII 125 Warranty Statement...

Страница 126: ...Warranty Statement This page intentionally left blank 126 Nio MKII This page intentionally left blank...

Страница 127: ...er and shall advise the Customer on how to proceed Customer Support may ask to return the faulty equipment or faulty subassemblies to the BarcoView Customer Support Center or an authorized service cen...

Страница 128: ...nance 2 5 Any failures resulting from an accident negligence such as but not limited to removing or deleting system files licensed software product files misuse circuit failure or any change damage du...

Страница 129: ...all implied warranties of merchantability and fitness for a particular purpose ARTICLE 6 LIMITATION OF LIABILITY Barcoview shall not under any circumstances be liable to customer or any third party f...

Страница 130: ...s and hinder performance by one party of his obligations ARTICLE 8 GENERAL 8 1 Customer acknowledges its understanding that all software and electronic devices including BarcoView products are subject...

Страница 131: ...www barco com B4100575 01 October 2006...

Отзывы: