R5915130 /05
Nio 2MP / 3MP LED Display
32
2.
Navigate to the
Embedded QA
/
Calibration Status
menu.
3.
Enter the
Ambient Light Compensation
submenu.
4.
Select
Enabled/Disabled
as desired and confirm.
4.14 Reading room
About reading room
Reading rooms can only be selected on your display when the display function is set to DICOM.
Please refer to
The American Association of Physicists in Medicine (AAPM) composed a list of pre-defined reading rooms.
Each of these reading rooms are defined by following parameters:
•
the maximum light allowed in this type of room
•
the preset ambient light correction value for this reading room
These parameters are stored in your display and determine the preset ambient light correction value to take
into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is enabled.
Please refer to
“Ambient Light Compensation (ALC)”, page 31
to enable ALC.
The available reading rooms for your Nio 2MP / 3MP LED Display are:
•
CR/DR/ MAMMO:
Corresponds to light conditions in diagnostic reading rooms for computed radiology,
digital radiology or mammography. This setting has the lowest maximum ambient light.
•
CT/MR/NM:
Corresponds to light conditions in diagnostic reading rooms for computed tomography,
magnetic resonance or nuclear medicine scans.
•
Staff Office:
Corresponds to light conditions in office rooms.
•
Clinical Viewing Room:
Corresponds to light conditions in diagnostic reading rooms for clinical viewing.
•
Emergency Room:
Corresponds to light conditions in emergency rooms.
•
Operating Room:
Corresponds to light conditions in operating rooms. This setting has the highest
maximum ambient light.
To select a reading room
1.
Bring up the OSD main menu.
2.
Navigate to the
Embedded QA
menu.
3.
Enter the
Reading Room
submenu.
4.
Select a desired reading room and confirm.
4.15 Embedded QA
About Embedded QA
Embedded QA allows you to run a display calibration or compliance test directly from the display using the
OSD menus described in the next sections. Embedded QA will use the front sensor / I-Guard to measure the
necessary luminance levels for either a calibration or compliance test. Various settings for both actions can be
selected from the display’s OSD menu. The last results of both actions can be consulted from the OSD.
Embedded QA or QAWeb Enterprise?
Embedded QA is not a replacement for the Barco QAWeb Enterprise solution.
Although Embedded QA is a reliable option to perform a simple calibration or compliance test, Barco still
highly recommends QAWeb Enterprise as the solution of choice for calibration and QA. QAWeb Enterprise
brings many benefits such as centralized asset management, the ability to schedule tasks, remote
management, automated reporting, alerting and specific support of regional QA standards such as DIN 6868-
57, JESRA and AAPM TG18. That's why QAWeb Enterprise Agent acts as the master for all supported
Содержание Nio MDNC-3521 SNES
Страница 20: ...R5915130 05 Nio 2MP 3MP LED Display 20 Installation and setup...
Страница 37: ...37 R5915130 05 Nio 2MP 3MP LED Display Cleaning the display 5...
Страница 39: ...39 R5915130 05 Nio 2MP 3MP LED Display Repackaging instructions 6...
Страница 44: ...R5915130 05 Nio 2MP 3MP LED Display 44 Repackaging instructions...
Страница 64: ...R5915130 05 Nio 2MP 3MP LED Display 64 Important information...
Страница 65: ......
Страница 66: ...www barco com R5915130 05 2023 08 22 Barco NV President Kennedypark 35 8500 Kortrijk Belgium...