Troubleshooting
19
Troubleshooting
The following problems may occur:
The sensor is connected, but the system does not find it
This may be the case when you unplugged the sensor, and
plugged back in while MediCal LE resided in the system tray.
Solve this by rebooting your computer.
Display window or test patterns appear dithered
Ensure that the selected color palette in the display properties is
set to True Color.
Solve this by selecting a True Color palette.
Patterns or calibration does not appear on all displays of the system
•
You may have added displays to your system while the
MediCal LE window was still open.
Close the application window and restart MediCal LE.
•
You may have deselected displays in the Options section.
Patterns or calibration has disappeared
•
Probably you have one or more imaging boards in your PC to
which no display is connected, while the board is still
activated in the display properties control panel.
Another possibility is that you have a multi-head imaging
board in the PC, while one of the heads is not connected to a
display.
De-activate these imaging board heads in the control panel
or connect a display to them.
•
You may have deselected displays in the Options section.