Calibrating displays
13
Calibrating displays
Note:
In the Options section, you can select which displays in the
system need or need not to be controlled. Calibration is possible
only on selected displays.
At least one display must be selected.
To calibrate displays:
1. Be sure to connect the optical sensor to the PC USB connector
or a USB hub.
2. The Found New Hardware wizard may appear, asking you to
search for a suitable driver.
You can find the sensor drivers on the MediCal LE CD-ROM.
For more information, see § “
“Optical sensor connection”
”.
3. When done, open MediCal LE.
4. If necessary, click on the Options button to:
• exclude displays from calibration
• select the display function to which to calibrate the
displays
When done, click on Back to return to the main window.
For more information, see §
“Options”
.
5. Click on the Calibrate button.
6. If you have a system with more than 1 display connected, a
window appears asking if you wish to calibrate this display.
Select Yes and click on the Next button.
If you have just 1 display connected, the calibration starts
immediately (see the following step).