Performing quality checks with test patterns
9
To display a custom test pattern
1. Open MediCal LE.
2. Click on the Test Patterns... button.
3. Double-click on the Browse icon.
4. An explorer-like dialog appears. Browse to the location
where the desired image is stored, select the image, and
click Open.
The selected image will appear.
Pattern options
When you click with the right mouse button while the pattern is
displayed, the pattern options appear:
Inverted:
The imaging board Look-Up Tables are inverted (white becomes
black etc.)
Calibrated:
Disable this option to view the test pattern without the effect of
the calibration. In this way, you can compare the calibrated and
uncalibrated situations.
Enable the option again for normal, calibrated operation.
Quit:
Close the test pattern on this display.
On a system with multiple displays (multi-head system), the
pattern will then appear on the next display.
You can also close the test pattern by left-clicking on the pattern
or by hitting any key on the PC keyboard.