Options
17
Options
1. Open MediCal LE.
2. Click on the Options button.
3. Select the display function for all displays of the system.
We recommend to select DICOM in almost any medical
application. The DICOM curve, if calibrated, ensures that you
can distinguish the highest possible number of gray levels.
Select CIE if you need to comply to the DIN or IEC standard.
Select Gamma to calibrate to a gamma 2.2 function.
Click on the Next button.
4. Select the units in which you want all luminance values and
measurements to be expressed: cd/m² (Candela per square
meter) or fl (Footlambert).
Note: Cd/m² is a metric unit and is synonymous with "nit".
Click on the Next button.
5. On a multihead system, a graphical representation of the
system appears. The rectangles in the “Select displays to be
controlled” window represent the display controller heads in
the system and the displays connected to them.
By default, all the displays are selected (showing a check
mark in the corresponding rectangle). To deselect displays
(meaning they will not be calibrated nor controlled), click the
corresponding rectangles. As a result, a cross appears in
these rectangles.
6. If you want all displays from the system to have the same
black and white reference values after calibration, check the
Equalize after Calibration option (multihead systems only).
7. Click on the Finish button to enter the selected options.
To cancel any changes made to the options, click on the Back
button to return to the main window.