K5902094 (451920611693) /02
MDSC-8232
45
Mission critical applications
We strongly recommend there is a replacement monitor immediately available in mission critical applications.
Use of Electrical Surgical Knives
Provide as much distance as possible between the electrosurgical generator and other electronic equipment
(such as monitors). An activated electrosurgical generator may cause interference with them. The interference
can activate the OSD menu of the display and as such disrupt the functionality of the display.
Power connection
–
Equipment with external 24 VDC power supply
•
Power requirements: The equipment must be powered using the delivered medical approved 24 VDC (
) SELV power supply.
•
The medical approved DC (
) power supply must be powered by the AC mains voltage.
•
The power supply is specified as a part of the ME equipment or combination is specified as a ME system.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
•
The equipment should be installed near an easily accessible outlet.
•
The equipment is intended for continuous operation.
Transient over-voltage
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
•
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1
3rd. Ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
•
To maintain compliance with EMC Regulation, use only shielded interface cables for the connection to
peripheral devices.
Power cords:
•
Europe: H05VV-F or H05VVH2-F PVC cord with appropriate EU plug.
US and Canada:
“
hospital grade
”
cord-set has to be used, provided with instructions to indicate that
grounding reliability can be achieved only when the equipment is connected to an equivalent receptacle
marked hospital only or hospital grade. These instructions need to be marked either on the equip. or on a
tag on the power cord
•
Do not overload wall outlets and extension cords as this may result in fire or electric shock.
•
Mains lead protection: Power cords should be routed so that they are not likely to be walked upon or
pinched by items placed upon or against them, paying particular attention to cords at plugs and
receptacles.
•
The power supply cord should be replaced by the designated operator only at all time.
•
Use a power cord that matches the voltage of the power outlet, which has been approved and complies
with the safety standard of your particular country.
Grounding reliability
Grounding reliability can only be achieved when the equipment is connected to an equivalent receptacle.
Water and moisture
The equipment is IP20 compliant. The monitor front side only is IP45 compliant. The PSU only is IP22
compliant.
Moisture condensation
•
Do not use monitor under rapid temperature and humidity change condition or avoid cold air from air-
conditioning outlet directly.
•
Moisture may condense on the surface or inside of the unit, or create a mist residue inside the protection
plate, this is not a malfunction of the product itself, although it may cause damage to the monitor.
•
If condensation happens, let the monitor stand unplugged until there is no condensation.
Содержание MDSC-8232
Страница 1: ...ENABLING BRIGHT OUTCOMES MDSC 8232 LED MDSC 8232 MNA User Guide 31 UHD 4K surgical display MDSC 8232 ...
Страница 5: ...5 K5902094 451920611693 02 MDSC 8232 Welcome 1 ...
Страница 8: ...K5902094 451920611693 02 MDSC 8232 8 Welcome ...
Страница 9: ...9 K5902094 451920611693 02 MDSC 8232 Parts controls and connectors 2 ...
Страница 15: ...15 K5902094 451920611693 02 MDSC 8232 Display installation 3 ...
Страница 22: ...K5902094 451920611693 02 MDSC 8232 22 Display installation ...
Страница 23: ...23 K5902094 451920611693 02 MDSC 8232 Daily operation 4 ...
Страница 29: ...29 K5902094 451920611693 02 MDSC 8232 Advanced operation 5 ...
Страница 41: ...K5902094 451920611693 02 MDSC 8232 41 4 Enable Disable DVI output as desired and confirm Advanced operation ...
Страница 42: ...K5902094 451920611693 02 MDSC 8232 42 Advanced operation ...
Страница 43: ...43 K5902094 451920611693 02 MDSC 8232 Important information 6 ...
Страница 63: ...K5902094 451920611693 02 MDSC 8232 63 Important information ...