5. Advanced operation

The American Association of Physicists in Medicine (AAPM) composed a list of pre-de

fi

ned reading rooms.

Each of these reading rooms are de

fi

ned by following parameters:

the maximum light allowed in this type of room

the preset ambient light correction value for this reading room

These parameters are stored in your display and determine the preset ambient light correction value to
take into account to recalculate the DICOM display function when Ambient Light Compensation (ALC) is
enabled. Please refer to "Ambient Light Compensation (ALC)", page 29 to enable ALC.

The available reading rooms for your Mammo Tomosynthesis are:

CR/DR/ MAMMO

Corresponds to light conditions in diagnostic
reading rooms for computed radiology, digital
radiology or mammography. This setting has the
lowest maximum ambient light.

CT/MR/NM

Corresponds to light conditions in diagnostic
reading rooms for computed tomography, magnetic
resonance or nuclear medicine scans.

Staff Of

fi

ce

Corresponds to light conditions in of

fi

ce rooms.

Clinical Viewing Room

Corresponds to light conditions in diagnostic
reading rooms for clinical viewing.

Emergency Room

Corresponds to light conditions in emergency
rooms.

Operating Room

Corresponds to light conditions in operating rooms.
This setting has the highest maximum ambient light.

To select a reading room:

1. Bring up the OSD main menu.

2. Navigate to the

Con

fi

guration > Calibration > Ambient Light

menu.

3. Enter the

Reading Room

submenu.

4. Select one of the available reading rooms and con

fi

rm.

5.18 Continuous ALC

About Continuous ALC

Continuous ALC can only be selected when the DICOM display function is selected.
Therefore, please refer to "Display functions", page 28 
to correctly set the display func-
tion.

Enabling continuous ALC will continuously recalculate the DICOM display function taking the averaged
ambient light into account.

To select continuous ALC:

1. Bring up the OSD main menu.

2. Navigate to the

Con

fi

guration > Calibration > Ambient Light

menu.

3. Enter the

Continuous ALC

submenu.

4. Select

Enabled/Disabled

as desired and con

fi

rm.

30

K5902035

 

MAMMO

 

TOMOSYNTHESIS

  07/12/2017

Содержание MDMG-5221

Страница 1: ...Mammo Tomosynthesis User Guide MDMG 5221 K5902035 07 07 12 2017...

Страница 2: ...Barco nv President Kennedypark 35 8500 Kortrijk Belgium Phone 32 56 23 32 11 Fax 32 56 26 22 62 Support www barco com esupport Visit us at the web www barco com Printed in Belgium...

Страница 3: ...OSD menus 22 4 7 Navigating through the OSD menus 22 5 Advanced operation 23 5 1 OSD menu position 23 5 2 OSD menu language 23 5 3 OSD menu automatic close function 23 5 4 Power status indicator light...

Страница 4: ...7 6 1 Cleaning instructions 37 7 Important information 39 7 1 Safety information 39 7 2 Environmental information 40 7 3 Regulatory compliance information 42 7 4 EMC notice 43 7 5 Explanation of symbo...

Страница 5: ...COM P14 compli ance QAWeb automatic Quality Assurance testing and remote assessment of the diagnostic area PPU Per Pixel Uniformity technology DURALIGHT superb luminance levels with an exceptionally l...

Страница 6: ...s two external power supplies a keyboard light a film clip If you ordered a Barco display controller it s also in the box together with its accessories A dedicated user guide is available on the syste...

Страница 7: ...AND CONNECTORS 2 1 Display front view Overview 7 1 2 3 4 5 6 Image 2 1 Front view 1 I Luminate Left key 2 Right key 3 Menu key 4 Power on off key 5 Ambient light sensor 6 Power LEDs 7 Side downstream...

Страница 8: ...partment cover 2 Display release button 3 Mounting anchor Connector compartment cover open 4 5 6 7 8 9 10 Image 2 3 Rear view cover open 4 24 Vdc power input 5 5 Vdc 0 5 A power output 6 Display Port...

Страница 9: ...oot Overview 1 2 3 4 2 Image 2 4 Dual tilt and swivel foot 1 Cable cover 2 Left right arm cover 3 Foot lock clip behind cable cover 4 Display positioning sliders 2 4 Accessories Film clip Image 2 5 Fi...

Страница 10: ...connectors Keyboard light Image 2 6 Keyboard light The keyboard light can be mounted under the display to provide a workspace light that is not interfering with the display image 8 K5902035 MAMMO TOMO...

Страница 11: ...ices 3 1 Mounting the displays Using the dual tilt swivel foot 1 Unpack the dual tilt swivel foot and place it on a solid surface 2 Unpack the first display 1 2 3 Hold the display slightly angled with...

Страница 12: ...e with the VESA 100 mm standard WARNING Never move a display attached to an arm by pulling or pushing the display itself Instead make sure that the arm is equipped with a VESA approved handle and use...

Страница 13: ...nnect the display to a display controller by using the DVI or the DP connection The active input can be selected in the OSD menu 1 Remove the cable cover To remove this cover push the lock clips insid...

Страница 14: ...3 Display installation 1 2 3 Connect the DVI or DP of the display controller to the DVI or DP connector of the display 12 K5902035 MAMMO TOMOSYNTHESIS 07 12 2017...

Страница 15: ...the display 7 Plug the mains connector of the external DC power supply to a grounded power outlet by means of the supplied power cable 8 Re attach the connector compartment cover Position the cover w...

Страница 16: ...s To route the cables If the display is mounted on the system dual tilt swivel foot cables are best routed behind the arms and through the cable channel in the foot shaft as shown in the pictures belo...

Страница 17: ...depth of opposite screws 7 Re attach the arm covers 3 6 Mounting the film clip To mount the film clip The film clip can be mounted on the top left or right side of the display 1 Position the film cli...

Страница 18: ...p in the display bezel gap 4 Push firmly against the front part of the white plastic clip so that the clip pops into the display bezel gap 5 With the enclosed Allen key tighten both screws at the bott...

Страница 19: ...he computer connected to your display If you have properly installed your display and display controller the Windows start up messages will appear once the boot procedure is finished Your Mammo Tomosy...

Страница 20: ...3 Display installation 18 K5902035 MAMMO TOMOSYNTHESIS 07 12 2017...

Страница 21: ...usage A good screen saver displays moving content In case you are working with the same image or an application with static image elements for several hours continuously so that the screen saver is no...

Страница 22: ...s always lit and available for activation of the I Luminate feature Please refer to I Luminate page 21 for more information The key auto dim function can be disabled in the OSD menus Please refer to K...

Страница 23: ...For both film modes the position can be set to Left Center Right When touching the I Luminate key the display will enter the default I Luminate mode according to the current setting for approximately...

Страница 24: ...0 seconds the OSD will disappear again The OSD menu auto exit function can be disabled in the OSD menu Please refer to OSD menu automatic close function page 23 for detailed instructions on how to do...

Страница 25: ...avail able for the OSD menu of your Mammo Tomosynthesis To change the language of the OSD menu 1 Bring up the OSD main menu 2 Navigate to the Configuration User Interface Menu menu 3 Enter the Langua...

Страница 26: ...lights will dim again if no further actions are taken within the following 5 seconds However this behavior can be changed so that the key indicator lights are always on or always off To configure the...

Страница 27: ...the Eth over USB submenu 4 Select Enabled Disabled as desired and confirm 5 8 USB About USB By default USB will be enabled This will allow the connected PC to communicate directly over USB with the i...

Страница 28: ...ate to the Configuration Power Management menu 3 Enter the DPMS Mode submenu 4 Select Enabled Disabled as desired and confirm 5 10 Hibernate About hibernate Enabling hibernation will not only switch o...

Страница 29: ...lt the I Luminate rectangle comes up in the top center of the screen This position can be changed to top left or top right corner To set the I Luminate film position 1 Bring up the OSD main menu 2 Nav...

Страница 30: ...SD main menu 2 Navigate to the Configuration Calibration menu 3 Enter the Viewing Mode submenu 4 Select Diagnostic Text as desired and confirm 5 15 Display functions About display functions Native unc...

Страница 31: ...mbient Light Compensation ALC About ALC Ambient Light Compensation ALC can only be enabled on your display when the DI COM display function is selected Therefore please refer to Display functions page...

Страница 32: ...e scans Staff Office Corresponds to light conditions in office rooms Clinical Viewing Room Corresponds to light conditions in diagnostic reading rooms for clinical viewing Emergency Room Corresponds t...

Страница 33: ...e man agement automated reporting alerting and specific support of regional QA standards such as DIN 6868 57 JESRA and AAPM TG18 That s why MediCal QAWeb Agent acts as the master for all supported dis...

Страница 34: ...After measurement the DICOM status report is shown To start DICOM compliance check 1 Bring up the OSD main menu 2 Navigate to the Configuration Calibration Embedded QA menu 3 Select DICOM compliance c...

Страница 35: ...tiply each individual pixel to one or more adjacent pixels so that the size of the displayed image will be a multiple of the original image source video input signal Image scaling is only possible whe...

Страница 36: ...ge HDR standard usually used on color displays HDR2 HDR2 is a Barco specific mode featuring 10 bit video encoding following the High Dynamic Range HDR standard but allowing full refresh rate which is...

Страница 37: ...confirm 5 24 EDID timings About EDID timings Following EDID timings are available for your Mammo Tomosynthesis Refresh Rate Allows to manually select the refresh rate of the image source video input...

Страница 38: ...ent status of your display runtimes tem peratures etc the status of the connected image sources video encoding mode timings etc the current calibration status of your display display function luminanc...

Страница 39: ...hol solvents at higher concentration 5 Strong alkalis lye strong solvents Acid Detergents with fluoride Detergents with ammonia Detergents with abrasives Steel wool Sponge with abrasives Steel blades...

Страница 40: ...6 Cleaning your display 38 K5902035 MAMMO TOMOSYNTHESIS 07 12 2017...

Страница 41: ...patient care equipment Equipment primarily for use in a health care facility that is intended for use where contact with a patient is unlikely no applied part The equipment may not be used with life...

Страница 42: ...ther closed location heed the necessary space between the set and the sides of the cupboard Installation Place the device on a flat solid and stable surface that can support the weight of at least 3 d...

Страница 43: ...system the lamps in the device contain mercury Turkey RoHS compliance T rkiye Cumhuriyeti AEEE Y netmeli ine Uygundur Republic of Turkey In conformity with the WEEE Regulation RoHS Chinese Mainland Ro...

Страница 44: ...ormation Products EIP that are sold within Chinese Mainland must comply with the Electronic Information Products Pollution Control Labeling Standard of Chinese Mainland marked with the Environmental F...

Страница 45: ...tlet on a circuit different from that to which the receiver is connected Consult the dealer or an experienced radio TV technician for help Changes or modifications not expressly approved by the party...

Страница 46: ...fer from harmful interference of surrounding equipment the user is encouraged to try to correct the interference by one or more of the following mea sures Reorient or relocate the receiving antenna or...

Страница 47: ...e 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5s Mains power quality should by that of a typical commercial or hospital environment If the user of the Mammo...

Страница 48: ...the user of the Mammo Tomosynthesis can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF com munications equipment transmitters and the Mammo...

Страница 49: ...lowing symbols nonrestrictive list Indicates compliance with the Directive 93 42 EEC as Class I device Indicates compliance with the Directive 93 42 EEC as Class II device Indicates compliance with Pa...

Страница 50: ...ates the DisplayPort connectors on the device Indicates the manufacturing date Indicates the temperature limitations6 for the device to safely operate within specs Indicates the device serial no Warni...

Страница 51: ...s of Barco products are subject to change without notice Trademarks All trademarks and registered trademarks are property of their respective owners Copyright notice This document is copyrighted All r...

Страница 52: ...0 150 kHz 15 80 Hz Housing color Gray Video input signals DVI D Dual Link DisplayPort per display Video inout terminals NA USB ports 1 upstream endpoint 3 downstream USB standard 2 0 Power requirement...

Страница 53: ...784 Additional keyboard light B4100574 QA software MediCal QAWeb Units per pallet NA Pallet dimensions W x H NA Warranty 5 years Operating temperature 0 C to 35 C 20 C to 30 C within specs Storage tem...

Страница 54: ...oc umentation source code README file or otherwise You shall not remove or obscure or otherwise alter the respective copyrights You acknowledge living up to the conditions of each separate Open Source...

Отзывы: