5. Important information

Immunity test

IEC 60601

Test levels

Compliance level

Electromagnetic
environment –
guidance

Electrostatic discharge
(ESD)

IEC 61000-4-2

± 6kV contact

± 8kV air

± 6kV contact

± 8kV air

Floors should be wood,
concrete or ceramic tile.
If

fl

oors are covered with

synthetic material, the
relative humidity should
be at least 30%

Electrical fast
transient/burst

IEC 61000-4-4

± 2kV for power supply
lines

± 1kV for input/ output
lines

± 2kV for power supply
lines

± 1kV for input/ output
lines

Mains power quality
should be that of a typical
commercial or hospital
environment

Surge

IEC61000-4-5

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

± 1 kV line(s) to line(s)

± 2 kV line(s) to earth

Mains power quality
should be that of a typical
commercial or hospital
environment

Voltage dips, short
interruptions and voltage
variations on power
supply input lines

IEC 61000-4-11

< 5% U

T

1

(> 95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

< 5% U

T

(>95% dip in

U

T

) for 5s

< 5% U

T

(> 95% dip in

U

T

) for 0.5 cycle

40% U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

< 5% U

T

(>95% dip in

U

T

) for 5s

Mains power quality
should by that of a typical
commercial or hospital
environment. If the user
of the JAO15 / JAO18
requires continued
operation during power
mains interruptions, it
is recommended that
the JAO15 / JAO18
be powered from an
uninterruptible power
supply or a battery.

Power frequency (50/60
Hz) magnetic

fi

eld

IEC 61000-4-8

3 A/m

Not applicable

2

Power frequency
magnetic

fi

elds should

be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3 Vrms

150 kHz to 80 MHz

3 V/m

80 MHz to 2.5 GHz

3 V

3 V/m

Portable and mobile
RF communications
equipment should be
used no closer to any
part of the JAO15 /
JAO18, including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance

d = 1.2

P

d = 1.2

P

80 MHz to 800

MHz

d = 2.3

P

800 MHz to 2.5

Ghz

1. is the a.c. mains voltage prior to application of the test level.

2. JAO15 / JAO18 doesn’t contain susceptible components to magnetic fields

K5903250 JAO15 / JAO18 10/03/2016

23

Содержание JAO18

Страница 1: ...JAO15 JAO18 User Guide Smart Terminal for the patient bedside K5903250 01 10 03 2016...

Страница 2: ...Barco NV President Kennedypark 35 8500 Kortrijk Belgium Phone 32 56 23 32 11 Fax 32 56 26 22 62 Support www barco com en support Visit us at the web www barco com Printed in Belgium...

Страница 3: ...10 3 Operation 13 3 1 On Off switching 13 3 2 Service call 13 3 3 USB charging 14 3 4 HDMI input 14 4 Cleaning 15 4 1 Cleaning instructions 15 5 Important information 17 5 1 Safety information 17 5 2...

Страница 4: ...Table of contents 2 K5903250 JAO15 JAO18 10 03 2016...

Страница 5: ...web IPTV VoIP for patient entertainment 16 9 15 6 JAO15 or 18 5 JAO18 1366x768 Fanless LED Panel Computer Intel Celeron J1900 Quad Core Low Consumption CPU Flat Panel by Resistive JAO15 or PCAP JAO18...

Страница 6: ...rpose configurable power key Please refer to the SDK documentation for more information about the different power key configura tion options 4 Combined SCR and RFID reader Combo Smart Carder Reader SC...

Страница 7: ...connector 7 Barcode scanner optional on JAO18 not supported on JAO15 8 Rear IO area SC 19V Image 1 4 Functional earth blind fastener 19V DC power supply input connector 6P6C RJ11 service call SC conne...

Страница 8: ...1 Introduction 6 K5903250 JAO15 JAO18 10 03 2016...

Страница 9: ...ch the wire to the functional earth blind fastener Image 2 1 WARNING When you unplug the power cable always pull the plug never pull the cable Although the JAO15 and JAO18 have a very low standby powe...

Страница 10: ...nt CAUTION Prior to removing any service door make sure the unit s power is off and disconnected from the power sources to prevent electric shock or system damage To fully disengage the power to the u...

Страница 11: ...ck the volume before listening the JAOHandset and do not listen at high volume levels for long periods To switch the position of the JAOHandset L R The default mounting position of your JAOHandset is...

Страница 12: ...ite dot to white dot pin Image 2 8 3 Use the screws from the handset box to firmly attach the handset to your terminal Image 2 9 2 5 VESA mounting To mount the JAO15 JAO18 on a VESA solution Your JAO1...

Страница 13: ...e washers W a different screw length L should be selected Please respect the following rule to select an appropriate screw length Lmin T W 4 mm Lmax T W 5 5 mm T W Lmin Lmax Image 2 11 CAUTION Use onl...

Страница 14: ...2 Installation 12 K5903250 JAO15 JAO18 10 03 2016...

Страница 15: ...t and the status LED of the JAOHandset can be detected controlled on the RJ11 Service Call connector of your JAO15 JAO18 Please refer to the connector pinout below for more details The electrical sign...

Страница 16: ...m USB ports USB charging is not possible via the USB ports at the back of your terminal 3 4 HDMI input Overview The JAO15 and JAO18 are capable of playing High Definition video and audio of any HDMI c...

Страница 17: ...Detergents with fluoride Detergents with ammonia Detergents with abrasives Steel wool Sponge with abrasives Steel blades Cloth with steel thread CAUTION Take care not to damage or scratch the front gl...

Страница 18: ...4 Cleaning 16 K5903250 JAO15 JAO18 10 03 2016...

Страница 19: ...been dropped and damaged The equipment has obvious signs of breakage Installation mounting When installing the equipment in a cupboard or another closed location heed the necessary space between the s...

Страница 20: ...your particular country Grounding Optionally earth the equipment by connecting the functional earth blind fastener to a grounded outlet by means of a wire with at minimum AWG18 size according to natio...

Страница 21: ...formance of procedures other than those specified herein may result in hazardous radiation exposure Recycling For more information about recycling of this equipment please contact your local city offi...

Страница 22: ...y participating in separate collection of batteries you will help to ensure proper disposal and to prevent potential negative effects on the environment and human health Turkey RoHS compliance T rkiye...

Страница 23: ...to the following two conditions 1 this device may not cause interference and 2 this device must accept any interference including interference that may cause undesired operation of the device This Cla...

Страница 24: ...000 3 3 Complies The JAO15 JAO18 is suitable for use in all establishments including domestic establishments and those directly connected to the public low voltage power supply network that supplies b...

Страница 25: ...5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5s Mains power quality should by that of a typical commercial or hospital environment If the user of the...

Страница 26: ...omer of the user of the JAO15 JAO18 can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equip ment transmitters and the JAO15...

Страница 27: ...icates the device meets the requirements of the applicable EC directives Indicates compliance with Part 15 of the FCC rules Class A or Class B Indicates the device is approved according to the UL Reco...

Страница 28: ...r Indicates the manufacturing date Indicates the temperature limitations6 for the device to safely operate within specs Indicates the device serial number Indicates the device part number or catalogue...

Страница 29: ...or damaged if not handled carefully when being stored Indicates a device that needs to be protected from moisture when being stored Indicates the storage direction of the box The box must be transpor...

Страница 30: ...demarks and registered trademarks are property of their respective owners Copyright notice This document is copyrighted All rights are reserved Neither this document nor any part of it may be reproduc...

Страница 31: ...1 IEC60950 1 CB ANSI AAMI ES60601 1 CAN CSA C22 2 No 60601 1 Weight 4 2 kg Audio Integrated stereo speakers 2 x 2 Watt down firing Internal digital microphone Backlight LED High luminance 300 cd m2 t...

Страница 32: ...mory 1 slot DDR3L 1333 SO DIMM socket Support up to 8GB capacity Standard 2 GB installed Storage pressure 50 kPa to 106 kPA Operating temperature 0 C to 40 C Storage temperature 20 C to 60 C Operating...

Страница 33: ...50 1 CB ANSI AAMI ES60601 1 CAN CSA C22 2 No 60601 1 Weight 4 8 kg Audio Integrated stereo speakers 2 x 2 Watt down firing Internal digital microphone Backlight LED High luminance 300 cd m2 typical Ba...

Страница 34: ...front Audio User interface Capacitive 10 point Multi touch screen P CAP touch technology Note Resistive single touch optional available from OEM program Multi purpose capacitive touch keys in glass fr...

Страница 35: ...4 7 0 1 mm 64 1 0 1 mm 269 7 0 1 mm 267 0 0 1 mm 50 0 0 1 mm 32 1 0 1 mm Image 5 1 Dimension drawing JAO18 482 0 0 1 mm 66 0 0 1 mm 62 5 0 1 mm 165 7 0 1 mm 99 6 0 1 mm 87 8 0 1 mm 308 2 0 1 mm 302 0...

Отзывы: