5. Important information
Separation distance according to frequency of transmitter
Rated maximum output
power of transmitter
5
W
150kHz to 80MHz
d=1.2
√
P
80MHz to 800MHz
d=1.2
√
P
800MHz to 2.5GHz
d=2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is af-
fected by absorption and re
fl
ection form structures, object and people.
5.5
Explanation of symbols
Symbols on the device
On the device or power supply, you may
fi
nd the following symbols (nonrestrictive list):
Indicates the device meets the requirements of the applicable EC directives.
Indicates compliance with Part 15 of the FCC rules (Class A or Class B)
Indicates the device is approved according to the UL Recognition regulations
Indicates the device is approved according to the UL regulations for Canada
and US
Indicates the device is approved according to the UL Demko regulations
Indicates the device is approved according to the CCC regulations
Indicates the device is approved according to the VCCI regulations
5. For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter. Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
K5903250 JAO15 / JAO18 10/03/2016
25
Содержание JAO18
Страница 1: ...JAO15 JAO18 User Guide Smart Terminal for the patient bedside K5903250 01 10 03 2016...
Страница 4: ...Table of contents 2 K5903250 JAO15 JAO18 10 03 2016...
Страница 8: ...1 Introduction 6 K5903250 JAO15 JAO18 10 03 2016...
Страница 14: ...2 Installation 12 K5903250 JAO15 JAO18 10 03 2016...
Страница 18: ...4 Cleaning 16 K5903250 JAO15 JAO18 10 03 2016...