Operation
E-2320 C
53
2. In the Calibration dialog (see step no. 3. in the description of
the calibration with sensor), check the option
Visually
and
click the
OK
button.
The DICOM Optimizer dialog appears.
Figure 23: DICOM Optimizer dialog
3. Adjust the slider until there is almost no visible difference
between the background of the image above the slider and
the bitmap inside this image.
4. When done, click button
Set point 1 of 20
.
5. Repeat this procedure until the button is grayed. This
indicates the last point is set.
6. Click
OK
to finish.
To reset to the latest calibration:
1. If you have multiple displays connected (multi-head system),
select the display you wish to reset to the latest calibration.
2. Click the
Reset
button.