Operation
48
E-2320 C
•
To select another color temperature, select an item from the
Color Temperature
drop-down box. This function is available
for E-2320 C displays only.
After selecting another color temperature it may be
necessary to recalibrate the display. Therefore, check the
Calibration tab to see if calibration is necessary: In case the
display is still in calibrated position, the text in the calibration
tab will say ‘The selected display is calibrated ...’. In case the
display is no longer in calibrated position, the text will say
‘The selected display uses the factory settings ...’.
•
To adjust display luminance manually, adjust the
Luminance
slider. This function is not available for MFCD/MFGD 1218
and MFCD 1219 displays.
The luminance setting is saved when the system is shut
down.
If the slider is positioned above the CAL mark, the luminance
is in calibrated position, as determined during the latest
calibration. If the slider is not above the CAL mark, the
luminance is not in calibrated position.
Important notices about DICOM compliance:
•
If you want the system to be DICOM compliant, you must
select the DICOM display function and calibrate the display.
After calibration, the luminance must remain in calibrated
position to maintain DICOM compliance.
•
When you have used NioWatch to calibrate the display to the
DICOM display function, it is the NioWatch application that
ensures DICOM compliance. Be aware that if you use the
display without NioWatch afterwards, the display is no
longer DICOM compliant. In that case, please consult Barco
MIS service to reset the display function to DICOM.