Operation
E-2320 C
51
2. If you have multiple Nio displays of the same type connected
(multi-head system), consider if you wish to
equalize
the
display you calibrate to a reference display.
The first calibrated display in the system is the reference
display.
Equalization
means the luminance of the display you are
calibrating will be matched to the luminance of the reference
display.
If you wish to equalize the display, you must close the
Display Settings dialog and make sure the
Equalization
option in the
Application Settings
dialog is set. See
“Calibration tab” in the description of the Application
Settings dialog.
To calibrate the display using the Barco LCD sensor:
1. If you have multiple displays connected (multi-head system),
select the display you wish to calibrate by clicking on the
corresponding numbered icon in the Display Settings dialog.
2. Click the
Calibrate
button.
3. A message appears, showing the different calibration
options.
Figure 21: Calibration choices
4. Select “Using the BARCO LCD sensor” and click
OK
.
This option is grayed in case NioWatch does not find a
connected Barco LCD sensor.