7. Important information
Electromagnetic immunity
The Coronis Uniti is intended for use in the electromagnetic environment speci
fi
ed below. The customer
or the user of the Coronis Uniti should assure that it is used in such an environment.
Immunity test
IEC 60601
Test levels
Compliance level
Electromagnetic
environment –
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood,
concrete or ceramic tile.
If
fl
oors are covered with
synthetic material, the
relative humidity should
be at least 30%
Electrical fast
transient/burst
IEC 61000-4-4
± 2kV for power supply
lines
± 1kV for input/ output
lines
± 2kV for power supply
lines
± 1kV for input/ output
lines
Mains power quality
should be that of a typical
commercial or hospital
environment
Surge
IEC61000-4-5
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality
should be that of a typical
commercial or hospital
environment
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
< 5% U
T
1
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
< 5% U
T
(> 95% dip in
U
T
) for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
< 5% U
T
(>95% dip in
U
T
) for 5s
Mains power quality
should by that of a
typical commercial or
hospital environment. If
the user of the Coronis
Uniti requires continued
operation during power
mains interruptions, it is
recommended that the
Coronis Uniti be powered
from an uninterruptible
power supply or a battery.
Power frequency (50/60
Hz) magnetic
fi
eld
IEC 61000-4-8
3 A/m
Not applicable
2
Power frequency
magnetic
fi
elds should
be at levels characteristic
of a typical location in
a typical commercial or
hospital environment.
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
3 V
3 V/m
Portable and mobile
RF communications
equipment should be
used no closer to any
part of the Coronis
Uniti, including cables,
than the recommended
separation distance
calculated from the
equation applicable
to the frequency
of the transmitter.
Recommended
separation distance
d = 1.2
√
P
1. is the a.c. mains voltage prior to application of the test level.
2. Coronis Uniti doesn’t contain susceptible components to magnetic fields
52
K5902079 CORONIS UNITI 21/04/2015
Содержание Coronis Uniti MDMC-12133
Страница 1: ...Coronis Uniti User Guide MDMC 12133 K5902079 03 21 04 2015...
Страница 6: ...1 Welcome 4 K5902079 CORONIS UNITI 21 04 2015...
Страница 30: ...4 Daily operation 28 K5902079 CORONIS UNITI 21 04 2015...
Страница 46: ...5 Advanced operation 44 K5902079 CORONIS UNITI 21 04 2015...
Страница 48: ...6 Cleaning your display 46 K5902079 CORONIS UNITI 21 04 2015...