R5911667 /03
AMM215WTTP
24
7.1 Safety Information
General recommendations
•
Read the operating manual thoroughly and be familiar with the content, prior to using this equipment.
•
Do not place the monitor, or any other heavy object, on the power cord to prevent fire or electrical shock
from damage to the power cord.
•
Power cord is used as a disconnection device. To de-energize equipment, disconnect the power cord. ME
equipment should be placed easily powered off place.
•
Do not expose the monitor to dust to avoid fire or shock hazard.
•
Avoid placing the monitor against a bright background or where sunlight or other light sources may reflect
on the area of the monitor. Place the monitor just below eye level.
•
Handle the monitor with care.
•
Refrain from giving a shock to the monitor.
•
Do not scratch the screen, as the screen is fragile.
Preventive maintenance
Periodic maintenance inspections are essential to keep the monitor in optimum condition and ensure safe
operation.
With the monitor disconnected from the mains, perform the following periodic check:
•
Check the integrity of the power cord and inspect its routing, so that it is not under the risk of being
punched or cut.
•
Check the integrity of the protective earth connection.
•
Clean the area around the power plug. Dust and liquids may result in fire.
•
Clean the ventilation slot of the monitor. Dust can obstruct the air flow and cause temperature increase of
the electronics.
General recommendations:
•
Keep the monitor clean to prolong its operational lifetime.
•
LCD panel performance may deteriorate in the long term. Periodically check that it is correctly operating.
•
Periodically check the tightness of the VESA mount screws. If not sufficiently tight, the monitor may detach
from the arm, which may result in injury or equipment damage.
Equipment
•
Monitor with power supply is suitable for use in patient environment.
•
This equipment has been tested and found to comply with the limits for medical devices in IEC 60601-1-
2:2014.
•
Equipment with SIP/SOP connectors should either comply with IEC 60601-1 harmonized national standard
or the combination should be evaluated. Do not touch the patient with signal input or output connectors.
Power connection
•
Only use the proprietary power supply.
•
Electrical input rating: 12V DC, 3A
•
Only use a hospital grade power supply cord.
•
Grounding reliability can only be achieved when the equipment is connected to an equipment receptacle
labeled
“
Hospital Only
”
or
“
Hospital Grade
”
.
•
To avoid the risk of electric shock, this equipment must only be connected to a supply mains with protective
earth.
•
The equipment is intended for continuous operation.
Recommendation to Use More Than One Unit
As unforeseen situations can occasionally occur with the monitor, when the display is used please use a
safety control. It is strongly recommended the user uses more than one unit or prepares a spare unit in case
there is an emergency.
Содержание AMM215WTTP
Страница 1: ...ENABLING BRIGHT OUTCOMES User Guide LCD Color Display AMM215WTTP...
Страница 4: ...R5911667 03 AMM215WTTP 4...
Страница 8: ...R5911667 03 AMM215WTTP 8 Welcome...
Страница 11: ...11 R5911667 03 AMM215WTTP Image 2 1 Installation...
Страница 13: ...13 R5911667 03 AMM215WTTP Daily Operation 3...
Страница 15: ...15 R5911667 03 AMM215WTTP Advanced Operation 4...
Страница 19: ...19 R5911667 03 AMM215WTTP Cleaning your display 5...
Страница 21: ...21 R5911667 03 AMM215WTTP Troubleshooting 6...
Страница 23: ...23 R5911667 03 AMM215WTTP Important Information 7...
Страница 37: ......