Shenchen Pump EasyPump Series
- 19
Pharmed tube. If need to use other material tubing, please contact with us for
technical support.
5.2
Tube Pressure Gap Adjustable Function
Adjust the pressure adjusting knob to adjust the tube pressure gap, to
change the pump head tube output pressure and flow rate.
Pressure adjustment is stepless adjustment, adjust to any position within
the adjustment range can be used normally.
A total of 5 scale values are set as indicators in the adjustment range.
Scale 3 is the factory standard scale value, which can be used as the
initial position of the tube pressure gap adjustment during the use of the
pump head. The factory default scale vale 3 is aligned with the gap
adjustment indicator line, as in below picture:
Gap adjustment
scale
Gap adjustment
indicator line
Turn the pressure adjusting knob clockwise, the scale value will increase,
tube pressure gap will reduce, the output pressure of the pump head tube
Содержание EasyPump I