MS series   

I   

 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 

5.  When desired stirring is completed, press the power switch to 

OFF

 

  position. Remove container from the top plate.

 

 

     

  If you need to stir a high viscosity liquid, we recommend to set a 

  slower stirring speed. 

  If the stir bar is not running smoothly, move container slightly, the 

  stir bar should be in the center of the container. 

 

 

 

Stirring Time 

The MS series magnetic stirrer installed with a clock chip for managing 
the stirring time. 
1.  Press the 

Time

 key, the display shows 000 min. 

2.  Press the 

 / 

key to set a desired stirring time. 

3.  Press the 

Enter

 key, the timer begins countdown until it gets to 

  zero, the stirrer will stop running. 
 

     

 

 

To exit the setup without saving changes, press the 

Exit

 key. 

 

 

 

Heating 

This section is applicable to models MS300 and 400 only. 
1.  Connect the temperature probe to the jack socket on the rear panel 
  of stirrer. 
2.  Place the temperature probe into the liquid. 
 

 

 
 
 

 

3.  Press the 

° C

 key, the display shows 000

 

° C. 

 

     

 

 

4.  Press the 

 / 

key to set a desired temperature. 

5.  Press the 

Enter

 key to begin the heating. 

 

 

  The hot plate will not heat if the temperature probe is not attached. 

  While heating a liquid the temperature probe must be immersed, 

  otherwise the hot plate will heat continuously and will go into a

.

 

  fault condition. 

  During the heating, the Heating icon will successively show as   

  follows: 

 

 

 

 

  When the liquid reached the target temperature, the Heating icon 

  will continuously flash. 

 

 

    

  Press the 

Exit

 key, the hot plate will stop working. 

 

 

 

Temperature Calibration 

During the heating, if the temperature reading displayed differs from 
that of an accurate thermometer, please follow the steps below to   
calibrate the temperature probe. 
1.  Turn off the stirrer.   
2.  Press and hold the 

 and 

keys for 5 seconds, then turn on the 

  stirrer. 
3.  Press the 

 / 

key to modify the temperature value. 

4.  Press the 

Enter

 key to save. 