background image

DOscan10 

 

Temperature Calibration 

1.    Press and hold the 

Cal

 key to enter the setup menu. 

2. 

Press the 

Enter

 key until the display shows

 

/

 

 

or

 

/

 

3. 

Press the 

Cal

 key, the tester enters the temperature calibration mode. 

4. 

Place the electrode into a solution with a known accurate temperature and wait for measurement is stable.   

5. 

Press the 

Cal

 key to modify the temperature value. 

6. 

Press the 

Enter

 key to save and press the 

Meas

 key to return to the measurement mode. 

 

 

 

 

 

Measurement 

The DOscan10 tester can be used to measure the water, wastewater, brine and other liquids. If your sample is seawater or water containing large 
amounts of salt, make sure to set the salinity coefficient before measurement. Some gas and steam such as chloride, sulfur dioxide, sulfureted 
hydrogen and carbon dioxide can permeate the membrane via diffusion. Their existence will influence the measurements. If the sample contains 
solvent, grease, sulfide and alga, the membrane will be damaged or eroded. 
 
1. 

Rinse the electrode with deionized water.   

2. 

Place the electrode into the sample solution and stir gently. Wait for the measurement to stabilize and record the reading. 

 

   

 

If the display shows "----" indicating the measurement exceeds the range, remove the electrode from the sample immediately. 

 

If the

 

 

option is enabled in the setup menu, the tester will automatically lock a measurement endpoint and show 

  icon. 

 

Press the 

Meas

 key to resume measuring. 

 

Electrode Maintenance 

 

Rinse the electrode thoroughly with deionized water after use. 

 

DO NOT touch the membrane and always keep it is clean and wet. 

 

If you do not use the electrode for long periods, screw off the membrane cap and rinse the electrode anode/cathode and membrane with 

 

deionized water and blot dry. Install the electrode and store dry. 

 

 

 

cathode 
 

anode 

DO membrane 

Содержание D0scan10

Страница 1: ...DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual Bante Instruments Inc...

Страница 2: ...ess the key again to resume measurement Exit the calibration setting and return to the measurement mode Start calibration Press and hold the key to enter the setup menu Select an option Confirm the ca...

Страница 3: ...to polarize Setup Menu The DOscan10 tester contains 8 menu items in the setup menu the following table describes the functions of each option Menu Description Options Description Default Set the sali...

Страница 4: ...enabled all of the calibration data and user specific settings will be deleted or reset the tester must be recalibrated The following table describes the relationship between the altitude and barometr...

Страница 5: ...0 relative humidity or place the electrode into the air saturated water for about 10 minutes Press the Enter key to begin the calibration 1 3 When the reading has stabilized the tester will show End a...

Страница 6: ...meate the membrane via diffusion Their existence will influence the measurements If the sample contains solvent grease sulfide and alga the membrane will be damaged or eroded 1 Rinse the electrode wit...

Страница 7: ...DO ES Electrolyte solution 480ml Specifications Dissolved Oxygen Model DOscan10 Range 0 0 to 20 0mg L Resolution 0 1mg L Accuracy 0 5mg L Saturation Range 0 0 to 200 0 Resolution 0 1 Accuracy 2 0 Othe...

Страница 8: ...warranty does not cover the electrode and membrane cap Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or in...

Отзывы: