DOscan10 

 

Filling the Electrolyte Solution 

1. 

Unscrew the membrane cap from the buttom of the electrode, rinse the inside and outside with deionized water and blot dry. 

2.      Fill the membrane cap halfway with electrolyte solution.   
3. 

Screw the membrane cap back onto the electrode. Some electrolyte solution will overflow during this process. 

4. 

Check the electrode, make sure that no air bubbles are trapped in the electrolyte solution and the membrane is not creased or damaged. 

 

 

 

 

Switching the Tester On and Off 

 

Press and hold the 

Meas

 key for about 5 seconds to switch on the tester. 

• 

  Press and hold the 

Meas

 key to switch off the tester. 

 

Polarizing the Electrode 

Connect the electrode to tester, switch on the tester and wait 10 minutes for the electrode to polarize. 
 

Setup Menu 

The DOscan10 tester contains 8 menu items in the setup menu, the following table describes the functions of each option. 
 

Menu 

Description 

Options 

Description 

Default 

 

Set the salinity compensation coefficient of sample 

 

Range: 0 to 35

 

g/L 

0

 

g/L 

 

Set the barometric pressure coefficient according to the 
local altitude 

 

Range: 450 to 850

 

mmHg 

760

 

mmHg 

 

Set the number of calibration points 

 

1 or 2 points 

1 point 

 

 

Set the measurement unit 

mg/L 

Milligrams per liter 

mg/L 

ppm 

Parts per million 

Percentage saturation 

 

Degrees Celsius 

° C 

 

Degrees Fahrenheit 

 

Calibrate the temperature 

 

Reading ± 10° C/° F 

--- 

 

 

When the option is enabled, the tester will automatically   
sense a stable reading and lock the measurement 

 

Enable 

Disable 

 

Disable 

 

When the option is enabled, the tester will automatically   
switch off if no key is pressed within 8 minutes 

 

Enable 

Disable 

 

Disable 

 

Reset the tester to factory default settings 

 

Enable 

Disable 

 

Disable 

 

no air bubbles

 

Содержание D0scan10

Страница 1: ...DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual Bante Instruments Inc...

Страница 2: ...ess the key again to resume measurement Exit the calibration setting and return to the measurement mode Start calibration Press and hold the key to enter the setup menu Select an option Confirm the ca...

Страница 3: ...to polarize Setup Menu The DOscan10 tester contains 8 menu items in the setup menu the following table describes the functions of each option Menu Description Options Description Default Set the sali...

Страница 4: ...enabled all of the calibration data and user specific settings will be deleted or reset the tester must be recalibrated The following table describes the relationship between the altitude and barometr...

Страница 5: ...0 relative humidity or place the electrode into the air saturated water for about 10 minutes Press the Enter key to begin the calibration 1 3 When the reading has stabilized the tester will show End a...

Страница 6: ...meate the membrane via diffusion Their existence will influence the measurements If the sample contains solvent grease sulfide and alga the membrane will be damaged or eroded 1 Rinse the electrode wit...

Страница 7: ...DO ES Electrolyte solution 480ml Specifications Dissolved Oxygen Model DOscan10 Range 0 0 to 20 0mg L Resolution 0 1mg L Accuracy 0 5mg L Saturation Range 0 0 to 200 0 Resolution 0 1 Accuracy 2 0 Othe...

Страница 8: ...warranty does not cover the electrode and membrane cap Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or in...

Отзывы: