DOscan10 

 

Setting the Default Option 

1. 

In the measurement mode, press and hold the 

Cal

 key for 5 seconds to enter the setup menu. 

2. 

If necessary, press the 

Cal

 key again to select an option. 

3. 

Press the 

Enter

 key, the tester saves the current option and moves to the next menu item. 

4. 

Repeat the steps above until the tester returns to the measurement mode. 

 

 

 

   

 

During the setting process, press the 

Meas

 key, the tester will exit the setup menu and return to the measurement mode. 

 

If you do not need to calibrate the temperature, press the 

Enter

 key to skip the

 

/

 

 

or

 

/

 

 

option. 

 

The

 

 

option is used to restore the tester back to the factory default settings. If enabled, all of the calibration data and user-specific 

 

settings will be deleted or reset, the tester must be recalibrated. 

 

The following table describes the relationship between the altitude and barometric pressure. Make sure to set the compatible parameter 
before the calibration and measurement. 

 

Altitude (m) 

kPa 

mmHg 

Altitude (m) 

kPa 

mmHg 

101.3 

760 

1800 

80.9 

607 

100 

100.1 

750 

1900 

79.9 

599 

200 

98.8 

741 

2000 

78.9 

592 

300 

97.6 

732 

2100 

77.9 

584 

400 

96.4 

723 

2200 

76.9 

577 

500 

95.2 

714 

2300 

76.0 

570 

600 

94.0 

705 

2400 

75.0 

563 

700 

92.8 

696 

2500 

74.1 

556 

800 

91.7 

688 

2600 

73.2 

549 

900 

90.5 

679 

2700 

72.3 

542 

1000 

89.4 

671 

2800 

71.4 

536 

1100 

88.3 

662 

2900 

70.5 

529 

1200 

87.2 

654 

3000 

69.6 

522 

1300 

86.1 

646 

3100 

68.7 

515 

1400 

85.0 

638 

3200 

67.9 

509 

1500 

84.0 

630 

3300 

67.0 

502 

1600 

82.9 

622 

3400 

66.2 

496 

1700 

81.9 

614 

3500 

65.4 

490 

 
 
 
 

option 

menu 

Содержание D0scan10

Страница 1: ...DOscan10 Pocket Dissolved Oxygen Tester Instruction Manual Bante Instruments Inc...

Страница 2: ...ess the key again to resume measurement Exit the calibration setting and return to the measurement mode Start calibration Press and hold the key to enter the setup menu Select an option Confirm the ca...

Страница 3: ...to polarize Setup Menu The DOscan10 tester contains 8 menu items in the setup menu the following table describes the functions of each option Menu Description Options Description Default Set the sali...

Страница 4: ...enabled all of the calibration data and user specific settings will be deleted or reset the tester must be recalibrated The following table describes the relationship between the altitude and barometr...

Страница 5: ...0 relative humidity or place the electrode into the air saturated water for about 10 minutes Press the Enter key to begin the calibration 1 3 When the reading has stabilized the tester will show End a...

Страница 6: ...meate the membrane via diffusion Their existence will influence the measurements If the sample contains solvent grease sulfide and alga the membrane will be damaged or eroded 1 Rinse the electrode wit...

Страница 7: ...DO ES Electrolyte solution 480ml Specifications Dissolved Oxygen Model DOscan10 Range 0 0 to 20 0mg L Resolution 0 1mg L Accuracy 0 5mg L Saturation Range 0 0 to 200 0 Resolution 0 1 Accuracy 2 0 Othe...

Страница 8: ...warranty does not cover the electrode and membrane cap Out of warranty products will be repaired on a charged basis The warranty on your tester shall not apply to defects resulting from Improper or in...

Отзывы: