Bante520 Portable Conductivity Meter
11
Temperature Calibration
During the measurement process, if the temperature reading displayed differs from that of an accurate thermometer, the meter needs to be calibrated.
1.
Connect the temperature probe to the meter and place into a solution with a known accurate temperature.
2.
Press and hold the
o
C
key to enter the temperature setting mode.
3.
Press the
or
key to set the temperature value.
4.
Press the
Enter
key to confirm. Calibrating is completed.
During the setting process, press the
or
key once, the setting value will increase or decrease by 0.1. Press and hold the
or
key, the
setting value will increase or decrease by 1.
Conductivity Measurement
1.
Rinse the conductivity electrode thoroughly with distilled water.
2.
Place the electrode into the sample solution, stir the electrode gently.
3.
Record the measured value when the reading is stable.
Auto-Hold
The meter contains an Auto-Hold function. If enabled, the meter will automatically sense a stable reading and lock the measurements, the HOLD icon
appears on the display. If disabled, press the key, the meter will immediately lock the displayed value. Press the
Meas
key to resume measuring.