Bante520 Portable Conductivity Meter
9
1.4
Place the electrode (and temperature probe) into the conductivity standard solution, the meter
automatically shows current calibration standard (e.g., 1413µ S/cm).
1.5
Press the
Enter
key, the meter begins the calibration, the Calibration icon will continuously
flashing.
1.6
Wait for the reading to stabilize, the display automatically shows END. Single point calibration
is completed.
Multi-point calibration
2.1 Ensure that you have selected 2 or 3 points calibration in the setup menu.
2.2
Repeat steps 1.2 to 1.5
above, when the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with
second point calibration.
2.3
Rinse and place the electrode (and temperature probe) into the standard solution, the display automatically shows current calibration standard
(e.g.,12.88mS/cm).
2.4
Press the
Enter
key, the meter begins the calibration, the Calibration icon will continuously flashing.
2.5
Wait for the reading to stabilize, the display will show “CAL3”. The meter prompts you to continue with third point calibrati
on.
2.6
Repeat the steps 2.3 and 2.4 above until the meter returns to the measurement mode. Calibration is completed.
Exit the calibration
During the calibration process, if you want to exit the calibration, press the
Meas
key.