Bante904P Portable Conductivity/DO Meter
5
Filling the Electrolyte Solution
1.
Take out the electrolyte solution from the carrying case. Unscrew the membrane cap from the buttom of the dissolved oxygen probe.
2.
Fill the membrane cap halfway with electrolyte solution.
3.
Screw the membrane cap onto the probe, excess electrolyte solution will drain out.
4.
Be sure the cathode of probe makes contact with membrane cap, the electrolyte solution in membrane cap should be without an air bubble.
Prior to Use
Remove the protective cap from the bottom of the electrode.
•
Conductivity Electrode:
Soak the electrode for a few minutes in tap water to remove dirt and oil stains on the electrode.
•
Dissolved Oxygen Probe:
Connect the probe to meter, turn on the meter 10 to 15 minutes and wait for the probe to polarize.
Cathode