background image

Bante904P Portable Conductivity/DO Meter 

 

Keypad 

 

KEY 

FUNCTION 

 

    Switches the meter ON/OFF. 

    Locks the measured value, press the key again to resume measuring. 

    Exits the calibration or setting and returns to measurement. 

 

    Toggles between available measurement modes. 

    Sets the temperature (Press and hold the key for 3 seconds). 

 

    Starts calibration. 

    Enters the setup menu (Press and hold the key for 3 seconds). 

 

    Stores current reading to memory. 

    Increase value or scroll up through the menu item. 

 

    Views the calibration report or data logs. 

    Decrease value or scroll down through the menu item. 

 

    Confirms the calibration, settings or displayed options. 

    Turn on/off the backlight (Press and hold the key for 3 seconds). 

 

Connectors 

 

               

 

INDEX: 

 

NO.  CONNECTOR 

DESCRIPTION 

6-pin Connector 

Used for connecting the conductivity electrode or dissolved oxygen probe 

Phone Jack 

Used for connecting the temperature probe 

USB 

Used for connecting the USB cable 

 

 

 

 

 
 
 

Содержание 904P

Страница 1: ...Bante904P Portable Conductivity DO Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Страница 2: ...conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following list describes the standard components...

Страница 3: ...rement mode Stable icon Indicates the measuring value has stabilized Calibration mode icon Indicates the meter is in the calibration mode Hold icon Indicates the measuring value has been locked Setup...

Страница 4: ...ress and hold the key for 3 seconds Stores current reading to memory Increase value or scroll up through the menu item Views the calibration report or data logs Decrease value or scroll down through t...

Страница 5: ...e meter allows using the DC5V power adapter with USB cable for power supply NOTE take out the batteries Connecting the Sensors Take out the electrode from the carrying case Conductivity Electrode or D...

Страница 6: ...be excess electrolyte solution will drain out 4 Be sure the cathode of probe makes contact with membrane cap the electrolyte solution in membrane cap should be without an air bubble Prior to Use Remov...

Страница 7: ...MENU DESCRIPTION OPTIONS DESCRIPTION DEFAULT Cell Constant Set the cell constant to match connected electrode K 0 1 K 1 K 1 K 10 User Defined Temperature Coefficient Set the linear temperature compens...

Страница 8: ...mg L mmHg Degrees Fahrenheit mg L Milligrams per liter ppm Parts per million mmHg Pressure unit kPa Pressure unit GENERAL OPTIONS MENU DESCRIPTION OPTIONS DESCRIPTION DEFAULT Stability Criteria When...

Страница 9: ...er the setup menu and the or key to select the menu item e g CELL P 01 1 2 Press the Enter key the display shows an option 1 3 Press the or key to select the desired option 1 4 Press the Enter key to...

Страница 10: ...splay the meter is now switched to the automatic temperature compensation mode Manual Temperature Compensation If the meter does not detect a temperature probe the C icon will show on the display indi...

Страница 11: ...ration mode To ensure higher accuracy we recommend that you perform 3 points calibration or select a standard value close to the sample value you are measuring The meter will automatically detect thes...

Страница 12: ...ation in the setup menu 2 2 Repeat steps 1 2 to 1 6 above when the first calibration point is completed the display will show CAL2 The meter prompts you to continue with second point calibration 2 3 R...

Страница 13: ...3 seconds to enter the setup menu the display shows CELL P 01 3 3 Press the Enter key the meter enters the setting mode 3 4 Press the or key until the display shows User Cell press the Enter key to co...

Страница 14: ...ment mode the display shows LOC P 01 4 2 Press the or key until the display shows ELE P 02 4 3 Press the Enter key the display shows the last calibration date Format month day 4 4 Press the key the di...

Страница 15: ...n icon continuously flashing 1 5 Wait for the reading to stabilize the meter automatically shows END Calibration is completed 2 points calibration 2 1 Ensure that you have selected 2 point calibration...

Страница 16: ...ompleted 2 points calibration 2 1 Ensure that you have selected 2 point calibration in the setup menu 2 2 Press the Cal key the meter shows 8 25mg L CAL 1 25 C press the or key until the display shows...

Страница 17: ...probe to the meter and place into a solution with a known accurate temperature 2 Press and hold the oC key for 3 seconds to enter the temperature setting mode 3 Press the or key to set the temperature...

Страница 18: ...e and other liquids If the sample is belong to the seawater or other water containing large amounts of salt please setting the salinity coefficient before measurement Some gas and steam such as chlori...

Страница 19: ...the low concentration liquids please use the default coefficient 0 0 ppt 1 Press and hold the key for 3 seconds to enter the setup menu and the or key until the meter shows SAL P 03 2 Press the Enter...

Страница 20: ...ins an Auto Hold function If enabled the meter will automatically sense a stable reading and lock the measurements the HOLD icon appears on the display If disabled press the key the meter will immedia...

Страница 21: ...appears on the display Viewing stored readings 1 Press the MR key in the measurement mode the meter shows LOC P 01 Data Log 2 Press the Enter key the meter shows the serial number of the stored data...

Страница 22: ...ished Click the OK button to confirm Click the Receive button the stored data automatically transfer to computer Interval recording This function is used for recording the measuring value within the s...

Страница 23: ...e screw off membrane cap and rinse the cathode anode and membrane with deionized water then soak up residual water on them with filter pape Install the probe again Troubleshooting LCD DISPLAY CAUSE CO...

Страница 24: ...ence Temperature 20 C or 25 C TDS Range 0 10 00 100 0 1000ppm 10 00 100ppt Max 200ppt Accuracy 1 F S TDS Factor 0 1 1 0 Default 0 5 Salinity Range 0 00 80 00ppt 0 00 42 00psu Accuracy 1 F S Resolution...

Страница 25: ...ccator CALIBRATION SOLUTIONS REAGENT 84 S cm Accurately weigh out 42 35mg of KCL and dissolve in 1 litre deionised water 1413 S cm Accurately weigh out 745 9mg of KCL and dissolve in 1 litre deionised...

Страница 26: ...ve Addendum 3 How to calculate the TDS conversion factor To determine the TDS conversion factor use the formula below Factor Where Actual TDS value from the high purity water and precisely weighed NaC...

Страница 27: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Отзывы: