background image

Bante904P Portable Conductivity/DO Meter 

11 

 

 

     
1.4 

Place the electrode into the conductivity standard solution, the meter automatically shows 

 

current calibration standard (e.g., 1413µ S/cm). 

1.5 

If necessary, press the 

 or 

 key to set the calibration value. 

 

 

   
1.6 

Press the 

Enter

 key, the Calibration icon begins flashing. 

 

   
1.7 

Wait for the reading to stabilize, the display automatically shows END. Single point calibration   

 

is completed. 

 

Multi-point calibration 

2.1    Ensure that you have selected 2 or 5 points calibration in the setup menu. 
2.2 

Repeat steps 1.2 to 1.6 

above, when the first calibration point is completed, the display will show “CAL2”. The meter prompts you to continue with 

second point calibration. 

2.3 

Rinse and place the electrode into the standard solution, the display automatically shows current calibration standard (e.g.,12.88mS/cm). 

2.4 

If necessary, press the 

 or 

 key to set the calibration value. 

2.5 

Press the 

Enter

 key, the Calibration icon begins flashing. 

2.6 

Wait for the reading to stabilize, the display will show “CAL3”. The met

er prompts you to continue with third point calibration. 

2.7 

Repeat the steps 2.3 and 2.5 above until the meter returns to the measurement mode. Calibration is completed. 

 
 
 
 
 

Содержание 904P

Страница 1: ...Bante904P Portable Conductivity DO Meter Instruction Manual BANTE INSTRUMENTS CO LTD...

Страница 2: ...conditions Relative humidity is less than 80 Ambient temperature is greater than 0 C and less than 60 C No potential electromagnetic interference The following list describes the standard components...

Страница 3: ...rement mode Stable icon Indicates the measuring value has stabilized Calibration mode icon Indicates the meter is in the calibration mode Hold icon Indicates the measuring value has been locked Setup...

Страница 4: ...ress and hold the key for 3 seconds Stores current reading to memory Increase value or scroll up through the menu item Views the calibration report or data logs Decrease value or scroll down through t...

Страница 5: ...e meter allows using the DC5V power adapter with USB cable for power supply NOTE take out the batteries Connecting the Sensors Take out the electrode from the carrying case Conductivity Electrode or D...

Страница 6: ...be excess electrolyte solution will drain out 4 Be sure the cathode of probe makes contact with membrane cap the electrolyte solution in membrane cap should be without an air bubble Prior to Use Remov...

Страница 7: ...MENU DESCRIPTION OPTIONS DESCRIPTION DEFAULT Cell Constant Set the cell constant to match connected electrode K 0 1 K 1 K 1 K 10 User Defined Temperature Coefficient Set the linear temperature compens...

Страница 8: ...mg L mmHg Degrees Fahrenheit mg L Milligrams per liter ppm Parts per million mmHg Pressure unit kPa Pressure unit GENERAL OPTIONS MENU DESCRIPTION OPTIONS DESCRIPTION DEFAULT Stability Criteria When...

Страница 9: ...er the setup menu and the or key to select the menu item e g CELL P 01 1 2 Press the Enter key the display shows an option 1 3 Press the or key to select the desired option 1 4 Press the Enter key to...

Страница 10: ...splay the meter is now switched to the automatic temperature compensation mode Manual Temperature Compensation If the meter does not detect a temperature probe the C icon will show on the display indi...

Страница 11: ...ration mode To ensure higher accuracy we recommend that you perform 3 points calibration or select a standard value close to the sample value you are measuring The meter will automatically detect thes...

Страница 12: ...ation in the setup menu 2 2 Repeat steps 1 2 to 1 6 above when the first calibration point is completed the display will show CAL2 The meter prompts you to continue with second point calibration 2 3 R...

Страница 13: ...3 seconds to enter the setup menu the display shows CELL P 01 3 3 Press the Enter key the meter enters the setting mode 3 4 Press the or key until the display shows User Cell press the Enter key to co...

Страница 14: ...ment mode the display shows LOC P 01 4 2 Press the or key until the display shows ELE P 02 4 3 Press the Enter key the display shows the last calibration date Format month day 4 4 Press the key the di...

Страница 15: ...n icon continuously flashing 1 5 Wait for the reading to stabilize the meter automatically shows END Calibration is completed 2 points calibration 2 1 Ensure that you have selected 2 point calibration...

Страница 16: ...ompleted 2 points calibration 2 1 Ensure that you have selected 2 point calibration in the setup menu 2 2 Press the Cal key the meter shows 8 25mg L CAL 1 25 C press the or key until the display shows...

Страница 17: ...probe to the meter and place into a solution with a known accurate temperature 2 Press and hold the oC key for 3 seconds to enter the temperature setting mode 3 Press the or key to set the temperature...

Страница 18: ...e and other liquids If the sample is belong to the seawater or other water containing large amounts of salt please setting the salinity coefficient before measurement Some gas and steam such as chlori...

Страница 19: ...the low concentration liquids please use the default coefficient 0 0 ppt 1 Press and hold the key for 3 seconds to enter the setup menu and the or key until the meter shows SAL P 03 2 Press the Enter...

Страница 20: ...ins an Auto Hold function If enabled the meter will automatically sense a stable reading and lock the measurements the HOLD icon appears on the display If disabled press the key the meter will immedia...

Страница 21: ...appears on the display Viewing stored readings 1 Press the MR key in the measurement mode the meter shows LOC P 01 Data Log 2 Press the Enter key the meter shows the serial number of the stored data...

Страница 22: ...ished Click the OK button to confirm Click the Receive button the stored data automatically transfer to computer Interval recording This function is used for recording the measuring value within the s...

Страница 23: ...e screw off membrane cap and rinse the cathode anode and membrane with deionized water then soak up residual water on them with filter pape Install the probe again Troubleshooting LCD DISPLAY CAUSE CO...

Страница 24: ...ence Temperature 20 C or 25 C TDS Range 0 10 00 100 0 1000ppm 10 00 100ppt Max 200ppt Accuracy 1 F S TDS Factor 0 1 1 0 Default 0 5 Salinity Range 0 00 80 00ppt 0 00 42 00psu Accuracy 1 F S Resolution...

Страница 25: ...ccator CALIBRATION SOLUTIONS REAGENT 84 S cm Accurately weigh out 42 35mg of KCL and dissolve in 1 litre deionised water 1413 S cm Accurately weigh out 745 9mg of KCL and dissolve in 1 litre deionised...

Страница 26: ...ve Addendum 3 How to calculate the TDS conversion factor To determine the TDS conversion factor use the formula below Factor Where Actual TDS value from the high purity water and precisely weighed NaC...

Страница 27: ...to speed up these efforts Warranty The warranty period for meter is one year from the date of shipment Above warranty does not cover the sensor and calibration solutions Out of warranty products will...

Отзывы: