SERVICE INFORMATION
The enclosed schematic contains complete voltage chart and all key wave
forms to assist in trouble shooting the Television Analyst.
The instrument contains two fuses. Both are accessible from the top of
the instrument and are located just in front of the 5BKPV1 scanning tube.
The Line Fuse is a 3 ampere fuse while the B+ fuse is a ¼ ampere slo-blo
fuse. Replacements must be made with identical types in order to retain the
protection offered by these fuses.
All tubes are accessible from the top of the Analyst and can be replaced
without removing the cabinet from the chassis.
All under the chassis components are available by removing the four
rubber feet and the two screws holding the perforated bottom plate to the
chassis.
If it becomes necessary to remove the cabinet perform the following steps.
1. Remove the three screws at the top of the panel.
2. Remove the screw at the right hand side of the panel located half
way between the top and the bottom of the panel.
3. Remove the screw at the left hand side of the panel located half
way between the top and bottom of the panel. DO NOT REMOVE
ANY ADDITIONAL SCREWS FROM THE PANEL.
4. Remove two screws from the right hand side of the case near the
bottom of the Analyst.
5. Remove two screws from the left hand side of the case near the
bottom of the Analyst.
6. Remove three screws from the rear of the cabinet located near the
bottom.
7. Lift off the case.
..