Maintenance and cleaning
Regular cleaning and care of the
fi
lter aggregate will ensure an optimum lifespan.
CAUTION!
Disconnect the
fi
lter aggregate from mains power each time before cleaning or care.
Do not hose down and do not expose to splashed water.
No guarantee is provided for damage caused by the use of unsuitable disinfection
agents or detergents.
5.1
Cleaning and care
• Use only commercially available care products without scouring additives - do not
use abrasive cleaners.
• Wipe down using only a moist cloth and then allow to dry or wipe dry.
5.2 Storage
If not used for longer periods of time, empty the
fi
lter aggregate completely and store it
covered in a cool, dry place.
6
Maintenance and repair
6.1 Maintenance
• Keep
the
fi
lter aggregate clean and dry.
• Replace used cleaning agents in the ultrasonic bath and the
fi
lter aggregate - do not
refresh by topping up.
• Filter cartridges must be replaced regularly - at the latest when the pressure switch
switches off.
The frequency with which the
fi
lter will need to be changed depends largely on the
type and degree of cleaning tank soiling.
Filter replacement:
Close the 3-way ball valve on the ultrasonic bath before replacing the
fi
lter, as
otherwise it will be running dry.
Pre-
fi
lter: Remove, clean/rinse or replace, depending on the degree of soiling. Rinse
the
fi
lter housing out thoroughly before replacing or re-inserting the
fi
lter
cartridge.