4 Use
4.1
Instructions for use
Check hose connections
• Check all hose connections between the
fi
lter aggregate and the ultrasonic bath for
leaks, especially after a longer period of non-use.
• The liquid must not exceed the maximum temperature of 70 °C.
Caution
!
Sediment particles can build up within the pre-
fi
lter,
fi
lling and blocking the
fi
lter
housing completely, resulting in dry runs that can damage the pump. The pressure
switch (on the
fi
ne
fi
lter) does not monitor clogging of the pre-
fi
lter.
4.2
General use
Step 1:
Preparation
Open the 3-way ball valve on the connected ultrasonic bath.
Open the ball valve on the
fi
lter aggregate.
Step 2:
Switching on
Press the On button (green) to switch on the
fi
lter aggregate.
Use the ball valve on the
fi
lter aggregate to set the desired
fl
ow-through rate.
Note:
Reducing return
fl
ow speed can eliminate any whistling noises in the ultrasonic bath.
Step 3:
Switching off the
fi
lter aggregate
If the cleaning tank is soiled with particles that tend toward sedimentation, make sure to
rinse with fresh cleaning agent before switching off the
fi
lter aggregate.
Failure to do so can result in pump damage.
Press the Off button (red) to switch off the
fi
lter aggregate.
For safety reasons, if the
fi
lter aggregate is to remain switched off for a longer
period, close the 3-way ball valve on the connected ultrasonic bath.