• When using strongly acidic preparations, the hard chromium plating of the ball valve may
become corroded and the ball valve start to leak.
If the use of a strongly acidic cleaning agent cannot be avoided, the use of a stainless steel ball
valve is recommended.
• When using preparations, the safety instructions included in the product lea
fl
ets are to be
fundamentally adhered to.
• Replace used sonication
fl
uids, do not refresh by adding
fl
uids.
4.2
Degassing the
fl
uid
Freshly-
fi
lled bath
fl
uid or
fl
uid that has remained in the oscillating tank for a longer period of time
must be degassed prior to use. See also Section 4.5.1.
Remove basket and other accessories from the oscillating tank.
Turn on the ultrasonic bath.
For setting the degassing time (10 min) and starting the DEGAS function, see Section 3.3.