2 Preparation
Carefully unpack the ultrasonic bath and accessories and inspect them for completeness
or possible transportation damages. If any damages or defects are found, these are to be
immediately noti
fi
ed in writing to the transportation company and to the supplier.
Before start-up, the ultrasonic bath is to be left to stand at its operating location for 2 hours so that
it may adapt to the ambient conditions.
2.1
Scope of delivery
1 Ultrasonic bath – see delivery note
1
Ball valve with hose, packaged separately with sealing tape and assembly instructions
1
Beaker holder (for DT 510 F) /
2
Beaker holder (for DT 1028 F)
1
TICKOPUR TR 3 (250 ml)
1 User
instructions
Additional accessories according to order – see delivery note
2.2
Set-up / assembly
• Place the ultrasonic bath atop a
fi
rm, level and dry surface. In doing so
-
observe the maximum weight of the ultrasonic bath, including
fl
uid.
Net weight see Technical Data Section 1.4.
-
Do not block the air supply below the ultrasonic bath.
-
Guard against moisture and wetness – risk of electric shock.
• Install the ball valve, hose socket and hose, which are included in the delivery, pursuant to the
enclosed assembly instructions.
2.3
Start-up
Thoroughly rinse the ultrasonic bath's oscillating tank with water before its
fi
rst use.
Connect the ultrasonic bath up to the power supply ( grounded socket).
Conduct function test. To do so, switch on the ultrasonic bath, start the ultrasound brie
fl
y and
then stop it again (maximum 1 to 2 seconds). A hissing noise should be heard. Then switch
device off again.
It is recommended that a foil test be conducted as part of quality assurance prior to the
fi
rst
use, see appendix for information.
Place
accessories in the ultrasonic bath as necessary.