parameter using the arrow key . Editing mode is cancelled with the "SET" or "START" keys.
The following operating parameters can be set:
•
Default values for the relative amplitude or power [Ampl or Power]
•
Default values for the sonication time [Time set]
•
Default values for the pulsation of the ultrasound (turn-on and turn-off time) [pulse]
• Limit value for the temperature monitoring [...°C]
The operating parameters "Energy" and "Elapsed sonication time" cannot be edited.
3.2.1
Sonication time
[Time set]
A value of 9 hours, 59 minutes and 59 seconds (9:59:59) is the maximum sonication time that
can be set. If the set value is exceeded during ultrasound operation, the sonication will stop. The
parameter "non-stop" is equivalent to unlimited continuous operation until ultrasound operation is
ended by pressing the "START/STOP" key.
a) Setting continuous operation [non-stop]
SET
Prog 1 Ampl
10%
Time set
00:01:00
elapsed
00:00:00
Pulse 4.0s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
Time set
00:01:00
Prog 1 Ampl
10%
Time set
00:01:00
elapsed
00:00:00
Pulse 4.0s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
Time set
nonstop
b) Setting the sonication time
If a value of 59 minutes or seconds is exceeded when setting the time, or if a shortfall occurs when
setting back the time, the values for the set hours or minutes will change accordingly. Example:
SET
Prog 1 Ampl
10%
Time set
00:01:00
elapsed
00:00:00
Pulse 4.0s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
Time set
00:01:00
Prog 1 Ampl
10%
Time set
00:01:00
elapsed
00:00:00
Pulse 4.0s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
00:01
Prog 1* Ampl
10%
Time set
00:01:00
elapsed
00:00:00
Pulse 4.0s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
00:02
Prog 1* Ampl
10%
Time set
00:01:00
elapsed
00:00:00
Pulse 4.0s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
00:02:00
Prog 1* Ampl
10%
Time set
00:01:00
elapsed
00:00:00
Pulse 4.0s 4.0s
Stop at
0°C
---°C
Energy 0.000
kJ
00:02:15
or
START
STOP
back to stand-by mode
SET
proceed to next set value or operating parameter