Pulsation
level
Work step
(sec)
Idle step (sec)
1
0.1
0.9
2
0.2
0.8
3
0.3
0.7
4
0.4
0.6
5
0.5
0.5
6
0.6
0.4
7
0.7
0.3
8
0.8
0.2
9
0.9
0.1
-
1
0
4
Use
4.1
Notes on use
• Do not use any bent probes (⇒ unstable operation, loss of power).
• Do not touch the sonication vessels with the oscillating probe - probes and vessels could be
damaged.
• The recommended immersion depth for probes is 10 - 20 mm, to prevent the aspiration and
mixing-in of air. If mixed-in air is desired, the probe may be immersed just a few millimetres.
In the case of greater immersion depths and/or the sonication of liquids with high viscosities,
the probe is more heavily damped. As a result, it is possible that the pre-set amplitude (power)
may not be reached, especially with higher default settings (>50 %). The reason is that the HF
generator can no longer provide the required power.
• The immersion depth may not exceed 20 mm in the case of micro and tapered tips.
• When producing emulsions, the probe should be immersed to the level of the interface between
the liquids to be mixed.
• Do not use combustible solvents in open reaction vessels since the operating safety of the
homogeniser could be compromised.
• In order to shut down the device, disconnect it from the mains ( pull the plug).