BB-200 Weighing Indicator – Programming and User’s Manual
Copyright © 2023 – B and B Scales, Inc. All Rights Reserved.
Page 26
Menu Name
Description
Possible Values
Example / Notes
C18
RS-232 Port
Settings
Controls how data is sent from
the RS-232 Port.
0 = Disable RS-232 Port
1 = Continuous Data
Transmit for a remote
display
2 = Send data only when
PRINT buttons is pressed
(to connect to a printer)
3 = Command Mode so
data can be requested by
a PC
4 = Continuous Request
Mode from a PC
C19
Baud Rate
Sets the Baud Rate of the RS-232
Port
0 = 1200
1 = 2400
2 = 4800
3 = 9600
C20
Manual Zero
Range
Set the manual zero functions
Option:
0= disable manual zero
1=±1% max capacity
2=±2% max capacity
C21
Initial Zero
Range
Set the Initial Zero Range.
This is the setting you would
change if you are weighing grain
and do not want the indicator to
auto-zero when you turn it on.
0=
disable
initial zero
1=±1% max capacity
2=±2
% max capacity
5=±5
% max capacity
10=±10
% max capacity
Set to “0” to have the
indicator “remember”
what was on the scale
when it was turned off.
Use the other settings to
auto-zero when you’re
weighing cattle, for
example, and they leave
manure on the scale.