Menu Name
Description
Possible Values
Example / Notes
C22
Automatic Zero
Tracking Range
This function defines a dynamic
range that the indicator will
interpret as ZERO. For example,
this could be caused when
ambient temperatures change in
a laboratory environment.
0= disable zero tracking
0.5=±0.5d
1.0=±1.0d
2.0=±2.0d
3.0=±3.0d
4.0=±4.0d
5.0=±5.0d
Notes:
1.
d = division
2.
The zero tracking
range can not bigger
than manual zero
range.
C23
Automatic Zero
Tracking Time
Defines the period of time over
which Automatic Zero Tracking is
performed.
0= Disable zero tracking
time
1= 1 second
2= 2 seconds
3= 3 seconds
C24
Overload
Range
Sets the tolerance, in divisions,
before the indicator displays an
Overload condition.
00= disable overload
range
01d
~
99d
Note: d =division
C25
Negative
Display
Defines how far negative the
indicator can go before reaching
an error condition
0=-20d
10=10% max. capacit
y
20=20% max. capacity
50=50% max. capacity
100=100% max. capacity
C26
Standstill Time
Determines how fast the scale
locks in on a weight.
0 = fast
1 = medium
3 = slow
Set this to slow if you are
using Animal Hold or if
you have a lot of filtering
set