MAINTENANCE
In-line measuring cells:
• Dirty glass discs must be cleaned to insure a correct measurement.
• The TURBISWITCH GA… measuring cell is supplied with a key tool for loosening the compression disc.
• The glass disc is removed from the cell with the suction cup.
• Drain the measuring cell completely before performing any maintenance or cleaning work!
• Do not use any objects or cleaners that might damage the glass disc!
• Lime scale can be removed with available light limescale remover.
• The side of the glass disc marked with a dot must be installed that it faces the medium!
• Check the seals (O-rings) for damage and replace them if necessary before start-up!
• The frequency of cleaning routines depends on operating conditions.
Measuring cell
Glass disc
O-ring
Compression disc Emitter or receiver
The side with the dot has an anti-fouling coating. Install this side that it faces the medium!