SETTINGS THE THRESHOLD, LIMIT VALUE AND COMMISSIONING (Continuation)

Other settings and optimisation:

• If the ascertained limit value is below 10% in the HIGH or MEDIUM range, turn anticlockwise (-) to switch into the LOW range, thus

permitting a more accurate adjustment.

• In order to avoid inadvertent switching due to any brief occurrence of gas bubbles or particles, switching contact delay time (TD) can be

increased to as long as 9.9 seconds.

• Inadvertent switching in the event of switching point fluctuation can be avoided by increasing reset hysteresis (HY).

TROUBLESHOOTING

Error

Cause

Remedy

The blue LED is always on.

The limit value is too low.

Set a higher ‘’Limit value’’!

Reset hysteresis is too high.

Set a lower ‘’Reset Hysteresis’’ value!

Turbidity exceeds the upper limit of the measuring range.

Check the measuring cell for any impurities

(Especially if the measurement previously

and remove them. Clean the glass

functioned in a permanently stable manner.)

discs!

“TT” appears on the display

Light from the emitter is not reaching the receiver.

Check/ Mount the emitter and/or receiver

or

to the measuring cell.

the HDR5 receiver blinks red.

Check the measuring cell for any impurities
and remove them.
Clean the glass discs.
The turbidity value is so high, that no light is
reaching the receiver.

HDR5 emitter blinks red.

Defective emitter.

Replace the emitter.

“TR” appears on the display.

The receiver has malfunctioned or it is not connected.

Check the connection of the receiver!

“ER” appears on the display.

Digital transmission between the receiver and the

Check the cable, observe EMC

GS5 is severely impaired.

environment, use a shielded cable.

“CR” appears on the display.

An internal memory error has occurred.

Reset to factory settings!

“HT” appears on the display.

The maximum permissible temperature of the

Comply with permissible temperature or

connected sensor has been exceeded.

switch off temperature monitoring in the
TP menu!

“LT” appears on the display.

The minimum permissible temperature of the

Comply with permissible temperature or

connected sensor has been fallen short of.

switch off temperature monitoring in the
TP menu!

The switching point has shifted.

Deposits on the optical windows.

Clean the glass discs!

The liquid level is too low.

Fill the measuring cell or immerse the
probe!

An incorrect turbidity range has been selected.

Change the turbidity range!

The limit value is too low or too high.

Adjust the limit value!

TUR

19-08-2021

M-410.03-EN-AA

410-03

/8

Turbidity Limit Value

Evaluation Unit

TURBISWITCH GS5

22, Rue de la Voie des Bans · Z.I. de la gare · 95100 ARGENTEUIL

Tel

Fax

+33 (0)1 30 25 83 20

+33 (0)1 34 10 16 05

Web

E-mail

www.bamo.eu

[email protected]

INTERNATIONAL

Содержание TURBISWITCH GS5

Страница 1: ...OMAT series TT GS TR GS and CP1 at the evaluation unit TURBISWITCH GS5 is not possible TECHNICAL DATA Supply power 100 240V AC 50 60Hz TURBISWITCH GS5 G or 10 30V DC and 12 24V AC TURBISWITCH GS5 D Po...

Страница 2: ...shed by simply turning the rotary pushbutton selector The selected hysteresis value is automatically skipped in this case 100 in the LOW range is followed by 0 in the MEDIUM range 100 in the MEDIUM ra...

Страница 3: ...iefly displayed The display is then limit value display INSTALLING A MEASURING CELL IN A PIPING SYSTEM IN LINE Installation of in line measuring cell The Measuring cells TURBISWITCH GA1 GA5 GA11 can b...

Страница 4: ...TURBISWITCH CP2 and CP5 transducers are designed as hanging probes or for use with an extension tube It can be installed in slurry tanks basins or open wells The probe should be installed so that it c...

Страница 5: ...eiver TT HDR and TR HDR or TT HDR5 and TR HDR5 Emitter Receiver IR Light TT HDR Emitter TURBISWITCH GS5 TR HDR Receiver 10 to 30 V DC 100 to 240 V AC L N Fuse 16 14 8 2 6 1 3 4 OC V RX V Limit value L...

Страница 6: ...13 15 Switching logic with increased reliability The contacts open when one of the following events occur Contact 9 11 opens when the limit value is exceeded Contact 9 12 opens when measurements are b...

Страница 7: ...NG THE THRESHOLD LIMIT VALUE AND COMMISSIONING Adjustment takes place after installation and electrical connection If problems arise see Troubleshooting All steps must be carried out in chronological...

Страница 8: ...is so high that no light is reaching the receiver HDR5 emitter blinks red Defective emitter Replace the emitter TR appears on the display The receiver has malfunctioned or it is not connected Check t...

Страница 9: ...mescale remover The side of the glass disc marked with a dot must be installed that it faces the medium Check the seals O rings for damage and replace them if necessary before start up The frequency o...

Отзывы: